To disconnect the Tester from the system make sure you leave the inlet adaptor on the instrument
and only disconnect at the end of the supply hose. The adaptor can then be disconnected afterwards
if required.
To turn off the Tester use the return arrows to return to the main menu screen and press the X in the
top right hand corner of the screen.
Note: - The unit will automatically turn off after a few minutes if left unused at the main menu screen.
6.0 Medical Gas Tests – Pharmacopeia
The "test complete" screen displays all the
measurements recorded during the test. And
automatically determines whether the test was a
pass or fail against the standard displayed in the
top right corner of the screen. The water content
is automatically also displayed, together with
oxygen content, ambient temperature and test
pressure. The test result is recorded in the
instrument's memory, which can store up to
20 previous tests.
Having selected the appropriate medical-gas
standard in the configuration screen, to carry out
a test press the "begin test" icon on the main
menu then follow the procedures as detailed in
5.0 Medical_Air Tests, with the following
exception.
When testing CO
(Hydrogen Sulphide) tube should be inserted in
the CO
port.
2
22
to EU Pharmacopeia the H
2
S
2