Operation; Dc + Ac Voltage Measurement; Dc Current Measurement - Velleman DVM902 User Manual

Cat ii 700 v / cat iv 600 v
Hide thumbs Also See for DVM902:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

10. Operation

10.1 DC + AC VOLTAGE MEASUREMENT

Do not measure circuits that may contain voltages > 700 VDC or
> 700 VAC
Use extreme caution when measuring voltages higher than 60 VDC
or 30 VAC rms.
Always place your fingers behind the protective edges of the test
probes while measuring!
1.
Set the rotary switch in the desired "
2.
Connect the red test lead to the "VHz" jack and the black lead to the
"COM" jack.
3.
Press SELECT to choose the "DC" or "AC" measurement.
4.
Connect the test leads to the source being measured.
5.
Read the voltage value and the polarity of the red test lead on the LCD
display.
6.
On the AC range, press "Hz/%" to measure frequency or duty cycle.

10.2 DC CURRENT MEASUREMENT

Do not measure circuits that may contain voltages > 700 VDC or
> 700 VAC
Use extreme caution when measuring voltages higher than 60 VDC
or 30 VAC rms.
Always place your fingers behind the protective edges of the test
probes while measuring!
1.
Set the rotary switch in the desired "
2.
Connect the red test lead to the "µAmA" jack and the black test lead to
the "COM" jack (switch the red lead to the "15A" jack for
measurements between 600 mA and 15 A).
3.
Press SELECT to choose the "DC" or "AC" measurement.
4.
Open the circuit in which the current is to be measured and connect
the test leads to the circuit IN SERIES.
5.
Read the current value and the polarity of the red lead connection on
the LCD display.
6.
On the AC range, press "Hz/%" to measure frequency or duty cycle.
V. 01 – 05/11/2019
DVM902
" or "
", "
13
" position.
" or "
" position.
©Velleman nv

Advertisement

Table of Contents
loading

Table of Contents