Duty Cycle; Operation; Voltage Measurement - Velleman DVM898 User Manual

6000 counts
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
999.9 kHz
9.999 MHz
Sensitivity: sine wave 0.6 V rms (9.999 MHz: 1.5 V rms)
Overload protection: 600 V DC or AC rms

9.9 DUTY CYCLE

0.1 %-99.9 %: ± (2.0 % rdg + 2 digits)
Frequency < 10 kHz
Sensitivity: sine wave 0.6 V rms
Overload protection: 600 V DC or AC rms
9.10 TEMPERATURE
range
°C/°F
NiCr-NiSi sensor
Overload protection: 600 V DC or AC rms

10. Operation

10.1 VOLTAGE MEASUREMENT

Do not measure circuits that may contain voltages > 600 VDC or
> 600 VAC
Use extreme caution when measuring voltages higher than 60 VDC
or 30 VAC rms.
Always place your fingers behind the protective edges of the test
probes while measuring!
1.
Set the rotary switch in the desired "
2.
Press "SELECT" to choose the AC or DC measurement mode.
3.
Connect the red test lead to the "VHz" jack and the black lead to the
"COM" jack.
4.
Connect the test leads to the source being measured.
5.
Read the voltage value and the polarity of the red test lead on the LCD
display.
Notes
In the AC range, press "Hz%" to measure the frequency or duty cycle.
V. 01 – 06/04/2017
100 Hz
1 kHz
resolution
1 °C/1 °F
DVM898
accuracy
-20 °C-150 °C
-4 °F-302 °F
151 °C-1000 °C
303 °F-1832 °F
" or "
12
± (3 °C + 1)
± (5 °F + 2)
± (3 % + 2)
± (3 % + 3)
"position.
©Velleman nv

Advertisement

Table of Contents
loading

Table of Contents