Correct Float Position - Fluke P3800C User Manual

High pressure hydraulic deadweight tester
Hide thumbs Also See for P3800C:
Table of Contents

Advertisement

1.888.610.7664
When reducing pressure ALWAYS use the capstan (by winding
out counter-clockwise). NEVER use any of the valves.
When reducing from high system pressures, some pressure is
still retained in the system, even after winding the capstan fully
out (approximately 700 psi / 50 bar). To release this pressure
see note below.
To reduce pressure to zero, wind the capstan fully out anticlockwise. Open
HP valve (1) SLOWLY and fully counter-clockwise. Open reservoir valve
(3), four turns anticlockwise.
Figure 3-8. Correct Float Position
www.calcert.com
15
10
20
5
25
0
30
W Caution
Note
ADJUST FLOAT HEIGHT OF
PISTON SO THAT THE
UNDERSIDE OF THE BOTTOM
WEIGHT LINES UP WITH THE
MID-FLOAT REFERENCE.
Operation
Generating Calibration Pressure
UPPER TRAVEL LIMIT
MID FLOAT POSITION
LOWER TRAVEL LIMIT
glf09.eps