Introduction / General Information; Description; Specifications - Instrutech VGC301A User Manual

Convection vacuum gauge controller
Table of Contents

Advertisement

Instruction Manual

1 Introduction / General Information

1.1 Description

The VGC301 vacuum gauge controller is a convenient and inexpensive power supply and readout instrument for
the InstruTech CVG101 Worker Bee™ convection-enhanced Pirani vacuum gauge or a Granville-Phillips
®
Convectron
gauge. The 1/8-DIN housing can be used as a bench top, or mounted in a cutout in an instrument
panel. The VGC301 is powered by 12-30 VDC, or by its self-contained universal 100-240VAC supply.
Thermal conductivity gauges measure pressure indirectly by sensing the loss of heat from a sensor to the
surrounding gases. The higher the pressure of the surrounding gas, the more heat is conducted away from the
sensor. Pirani thermal conductivity gauges maintain a sensor (usually a wire) at some constant temperature,
and measure the current or power required to maintain that temperature. A standard Pirani gauge has a useful
measuring range of about 10
above 1 Torr, convection-enhanced Pirani gauges increase the measuring range to just above atmosphere.
The VGC301 signals and relay functions are the same as found on similar convection gauge controllers from
other manufacturers. The VGC301 Controller, CVG101 vacuum gauge tube and gauge cable can be directly
interchanged with Brooks Automation, Inc. / Granville-Phillips
cable. (Remote interface, relay and power connectors are different.)

1.2 Specifications

measurement range (signal)
display
display engineering units
display update rate
weight
operating temperature
storage temperature
humidity
altitude
analog output (user-selectable)
serial communications
housing
input power
setpoint relays
connectors
RF/EMI protection
InstruTech, Inc.
-4
Torr to 10 Torr. By taking advantage of convection currents that are generated
-4
1.0 x 10
to 1000 Torr / 1.3 x 10
LCD, 4 digits, user-selectable Torr, mbar, or Pa,
(4 digits from 1100 Torr to 1000 Torr), (3 digits from 999 Torr to 10.0 mTorr),
(2 digits from 9.9 mTorr to 1.0 mTorr ), (1 digit from 0.9 mTorr to 0.1 mTorr)
Torr, mbar, or Pa - user selectable
0.5 sec
9 oz (250 g)
o
0 to +40
C
o
-40 to +70
C
0 to 95% Relative Humidity, non-condensing
operating
6,560 ft (2,000 m) max
storage
41,000 ft (12,500 m) max
a) non-linear S-curve 0.375 to 5.659 Vdc, or
b) linear 0 to 10 Vdc - user scalable, or
c) log-linear 1 to 8 Vdc, 1V/decade
RS232 and 2 wire/4 wire RS485 - ASCII protocol
1/8-DIN panel-mount enclosure
100-240 Vac, 50/60Hz, universal power supply or 12-30 Vdc, 2 W
two single-pole double-throw relays (SPDT), 1A at 30 Vdc resistive, or ac non-inductive
gauge: 9-pin D-sub female
analog output and digital interface: 9-pin D-sub male
relay outputs: 6-pin pluggable terminal block
power: universal AC and 2-pin pluggable terminal block DC
CE compliant
®
375 controller, 275 Convectron
-4
to 1333 mbar / 1.3 x 10
VGC301A Controller
®
275
®
gauge and gauge
-2
Pa to 133 kPa
Page 3