Using The Gauge With Different Gases - Instrutech VGC301A User Manual

Convection vacuum gauge controller
Table of Contents

Advertisement

Instruction Manual

5 Using the gauge with different gases

A thermal conductivity gauge senses heat loss, which depends on the thermal conductivity of the gas surrounding the
sensor. Since different gases, and mixtures, have different thermal conductivities, the indicated pressure readings and
outputs will also be different. InstruTech convection gauges (and most other thermal conductivity gauges) are normally
calibrated using nitrogen (N
thermal conductivity between nitrogen (N
different gases affect the display and output from an InstruTech convection gauge.
WARNING! Using a thermal conductivity gauge with gases other than that for which it is calibrated could result in
death or serious injury. Be sure to use gas correction data in this manual when measuring pressures of gases other than
N
/ air.
2
For N
the calibration shows excellent agreement between indicated and true pressure throughout the range from 10
2
1000 Torr. At pressures below 1 Torr, the calibration curves for the different gases are similar. The difference in readings at
these low pressures is a constant, a function of the difference between thermal conductivities of the gases.
At pressures above 1 Torr, indicated pressure readings may diverge significantly. At these higher pressures convection
currents in the gauge become the predominant cause of heat loss from the sensor and calibration depends on gauge tube
geometry and mounting position as well as gas properties.
Generally, air and N
are considered the same with respect to thermal conductivity, but even N
2
differences in readings at higher pressures. For example, when venting a system to atmosphere using N
readings change by 30 to 40 Torr after the chamber is opened and air gradually displaces the N
other gases the effect is much more significant and may result in a hazardous condition as described below.
Other considerations when using gases other than N
Flammable or explosive gases
WARNING! InstruTech convection gauges are not intrinsically safe or explosion proof and are not intended for use in
the presence of flammable or explosive gases or vapors.
Under normal conditions the voltages and currents in InstruTech convection gauges are too low to cause ignition of
flammable gases. However, under certain failure conditions, sufficient energy could be generated to cause flammable
vapors or gases to ignite or explode. Thermal conductivity gauges like the InstruTech convection gauges are not
recommended for use with flammable or explosive gases.
Moisture / water vapor
In some processes (lyophilization, for example) the gas composition may not change significantly, except for moisture
content. Water vapor can significantly change the response of a thermal gauge and correction should be made, as you
would for any other gas.
Other contaminants
If your gases condense, coat, or corrode the sensor, the gauge calibration and response to different gases will change.
Generally, if the gauge can be "calibrated" ("zero" and "span" settings), these changes are small enough to be ignored. If
you can't set zero and span, the gauge should be replaced or return to factory for evaluation and possible cleaning.
InstruTech, Inc.
). When a gas other than N
2
) and the gas in use. The charts and tables on the following pages indicate how
2
2
/ air is used, correction must be made for the difference in
2
/ air
VGC301A Controller
-4
to
and air will exhibit slight
2
, you may see
2
in the gauge. For most
2
Page 17