Agilent Technologies 87512A Operation And Service Manual page 26

Transmission/reflection test kit
Table of Contents

Advertisement

Equipment
Network Analyzer
Cable
1
Pad
Signal Generator
Power Meter
Power Sensor
50 Termination
This equipment is used for the insertion loss test with an 8751A.
1
This equipment is furnished with the 87512A/B.
2
11852B is furnished with 87512B.
3
This equipment is used for the insertion loss test without an 8751A.
4
4-2
Operational Verification
Table 4-1. Recommended Test Equipment
Critical Speci cations
No substitute
1
N(m)-N(m) Cable
Semi-rigid
Minimum Loss Pad, 50 - 75
Frequency: 500 MHz
4
Output Level: 0 dBm
Frequency: 500 MHz
4
Amplitude: -15.7 to 0 dBm
Frequency: 500 MHz
4
Amplitude: -15.7 to 0 dBm
Type: N-female
4
Recommended Model
8751A
PN 8120-4666
2
PN 41952-61601
2
11852B
3
8663A
436A
8482A
909C Opt. 013, 2 ea.