Reflection Coefficient/Return Loss
After performing a one port calibration, connect the test device to the TEST PORT. Then,
select the A/R or S
measurement function . The re
ection coecient is displayed by
11
selecting the LIN MAG display function. The return loss is displayed by selecting the LOG
MAG display function.
Complex Impedance
Complex impedance measurements are made by the analyzer's CONVERSION function to
convert the re
ection coecient to the impedance parameter. Refer to the analyzer's manual.
Two Port Measurements
Two Port Calibration
The 87512A/B should be calibrated for transmission measurements using the internal
calibration function of the analyzer. The calibration function requires the standards used to
be attached in place of a test device during calibration. Refer to the analyzer's manual for a
complete description of the calibration procedure.
Transmission Measurements
By using the 87512A/B transmission calibration and the ratio measurement functions, the
gain or loss of the device under test can be displayed directly in decibels or linear ratio units.
The calibration function removes the eects of the 87512A/B and the connecting cables. The
analyzer measurement functions B/R or S
calculates the gain or loss of the test device. The
21
test device is connected between the TEST PORT and the analyzer B port using the
exible
cable included in the test set.
Operation
3-3