Victron energy Lynx Smart BMS 500 Manual page 36

Table of Contents

Advertisement

VictronCo
GX device
nnect
warning
warning
code
code
W-D04
#224
VictronCo
GX device
nnect
alarm code
alarm code
A-B01
#103
#105 or
A-B02
#106
A-B06
-
VictronCo
GX device
nnect error
error code
code
E-B09
#09
E-B11
#11
E-B25
#25
E-B26
#26
E-B34
#34
E-B35
#35
E-B36
#36
E-B119
#119
GX device code
VictronConnect code
#201
#202
#203
#204
#205
#206
#207
#208
#209
#210
Page 33
Lynx Smart BMS
Message
Communication with distributor D
lost
Message
Low cell voltage
High current
Load disconnected
Message
Battery voltage not allowed
Hardware error
Pre-charge error
Contactor error
Wrong system voltage
Pre-charge timeout
ATC/ATD failure
Settings data lost
Message
A-F01
Fuse "Fuse_1" is blown
A-F02
Fuse "Fuse_2" is blown
A-F03
Fuse "Fuse_3" is blown
A-F04
Fuse "Fuse_4" is blown
A-F05
Fuse "Fuse_5" is blown
A-F06
Fuse "Fuse_6" is blown
A-F07
Fuse "Fuse_7" is blown
A-F08
Fuse "Fuse_8" is blown
A-F09
Fuse "Fuse_9" is blown
A-F10
Fuse "Fuse_10" is blown
Instructions / remarks
Check the cable between BMS and Distributor.
Instructions / remarks
Charge battery. The system will turn the loads back
on when the battery is sufficiently charged.
Reduce charging current or turn off some loads. The
system will try to re-enable chargers or loads in 5
minutes.
The loads have been turned off via the ATD contact
and/or via the GX device. Resolve this alarm by
charging the battery. If not resolved, eventually the
contactor will open and the DC system will be
disconnected.
Instructions / remarks
The battery voltage is too high or too low. Check the
battery voltage and check the battery settings in the
VictronConnect app.
This error occurs when the battery voltage is
outside all system voltage ranges (9V > Vbat > 60V)
Contact your Victron dealer.
The load resistance is too low to pre-charge the
loads. Disconnect or reduce some DC loads.
Contact your Victron dealer.
Check battery voltage settings in the
VictronConnect app.
The load capacity is too high to pre-charge.
Disconnect some DC loads.
Check ATC/ATD wiring and make sure that all loads
and chargers are controlled by ATC or ATD.
Settings data is corrupt. Go to the settings page and
reset to defaults.
Appendix

Advertisement

Table of Contents
loading

Table of Contents