Product Identification; Storage And Transport - Endress+Hauser Liquiphant FTL64 Brief Operating Instructions

Hide thumbs Also See for Liquiphant FTL64:
Table of Contents

Advertisement

Mounting
4.2

Product identification

The following options are available for the identification of the device:
• Nameplate specifications
• Extended order code with breakdown of the device features on the delivery note
• Enter serial number from nameplates in W@M Device Viewer
www.endress.com/deviceviewer. All of the information on the measuring device is
displayed along with an overview of the scope of technical documentation provided.
• Enter the serial number on the nameplate into the Endress+Hauser Operations app or scan
the 2-D matrix code on the nameplate with the Endress+Hauser Operations app
4.2.1
Electronic insert
Identify the electronic insert via the order code on the nameplate.
4.2.2
Nameplate
The information that is required by law and is relevant to the device is shown on the
nameplate.
4.3

Storage and transport

4.3.1
Storage conditions
Use original packaging.
Storage temperature
–40 to +80 °C (–40 to +176 °F)
Optional: –50 °C (–58 °F), –60 °C (–76 °F)
Transporting the device
• Transport the device to the measuring point in the original packaging
• Hold the device by the housing, temperature spacer, flange or extension pipe
Take suitable measures to protect the coating!
Do not bend, shorten or extend the tuning fork.
5
Mounting
WARNING
L
Loss of protection rating if the device is opened in a wet environment.
Only open the device in a dry environment!
Mounting instructions
• Any orientation for device with short pipe up to approx. 500 mm (19.7 in)
• Vertical orientation from above for device with long pipe
• Minimum distance between the fork tip and the tank wall or pipe wall: 10 mm (0.39 in)
6
Liquiphant FTL64
Endress+Hauser