12.9.2 Maintenance Checklist
Action
Verify Ambient Temperature, <32.2°C
* Adjust door handle for firm latching, as needed
Check and clean probe cover, gaskets, hinges and
door(s) of ice and snow.
See Figure 1-4 for probe location.
See "Cleaning the Door Gasket".
Check air filter. Clean or replace as needed.
See "Cleaning the Air Filter".
Check alarm back-up battery.
See "Connecting the unit to
Electrical Power" in Section 1 and
"Replacing the Battery" Refer
Check condenser fan motor for unusual motor noise
or vibration.
* Verify and document calibration, at the minimum,
annually. See Section 2 Calibration
* Clean condenser compartment and wipe off
condenser See "Cleaning the Condenser" in Section
4
* Qualified Service Technician only
** Dispose of properly, according to all state and federal regulations.
To minimize ice build-up inside of freezer:
•
Locate the freezer away from drafts or heating/cooling vents
•
Keep the number of door openings to a minimum
•
Minimize the length of time door is open
•
Make sure door latches securely after opening
Version 1.0
Monthly
User Manual
Yearly
√
√
√
More frequent cleaning may be required, depending
on use and environmental conditions.
√
√
** Replace
√
√
√
Every 2 Years
36