· The Marina basic / Marina deluxe cannot be used on a surface with a gradient of
over 10 degrees forwards / backwards.
· The Marina basic / Marina deluxe may only be used indoors.
· Don't touch the batteryterminals and the client at the same time.
· Portable and mobile RF communication equipment can affect medical devices.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the Marina deluxe / Marina basic electric including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
· Medical electrical devices involve special EMC precautions and must be used as per
the EMC information outlined in the instructions for use.
· The emissions characteristics of the Marina basic / Marina deluxe electric make it
suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in
a residential environment (for which CISPR 11 class B is normally required) this
equipment might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or
re-orienting the equipment.
· The trolley must not be used adjacent to or stacked on other products. However,
should that be necessitated, it must be checked whether the trolley functions
normally in this configuration too.
Maintenance
· If the charger or cable become damaged, they must be replaced (for electric
adjustable Marina basic / Marina deluxe.
· If the mattress is damaged, replace the mattress for hygiene reasons.
· It is recommended to replace the mattress after 5 years for hygiene reasons.
· The Marina basic / Marina deluxe must be serviced and inspected at least once a year.
· No modification of the Marina basic / Marina deluxe is allowed, except length
adjustment according to the instructions of Lopital.
· If the Marina basic / Marina deluxe is adjusted in length, appropriate inspection must
be conducted to ensure continued safe use of the shower trolley.
· Servicing and maintenance of the Marina basic / Marina deluxe may only be carried
out by persons with a technical background, who have experience in servicing and
maintaining similar products.
· Constructive components of the Marina basic / Marina deluxe may only be replaced
by suitably authorised personnel.