ARA Instrument User Manual V1.1
Attach the regulator, tubing and calibration cap to the cylinder so it's ready for use.
Click 'Sensor Calibration' (a new window will pop-up)
NOTE:
Ensure that the Span value is the same as the calibration gas concentration. If not use the Up/Down
arrows to adjust (refer to the label on calibration gas cylinder)
IMPORTANT:
Ensure you are in a clean fresh air environment.
Click Calibration, a zero is then established (very quickly)