PDM and Base Station/DoseView memories

Dose rate samples that have been overwritten in the PDM's dose rate
memory may still be available in the Base Station.
If there are no dose rate samples neither in the Base Station, nor in the
PDM dose rate memory, the Base Station and DoseView will instead display
mean dose rate values based on accumulated dose values (see figure below).
Lack of dose rate samples in the Base Station and/or DoseView memories
occurs when the:
• PDM is not within range of the Base Station when it is exposed to radiation.
• Dose rate samples in the PDM dose rate memory are overwritten.
Figure 10.1
dose rate values per hour calculated from the accumulated dose (blue curve).
N O T E
The same effect as illustrated in the figure above is also obtained by un-checking the
Show details check box in the Data filter panel (see section "View dose graph" on
page 4-12 ).
Document version 1.2
PDM and Base Station/DoseView memories
DoseView chart. In lack of dose rate samples, the yellow curve displays mean
10.5
Technical data
10-3