Technical data

PDM radiology

Table 10.1
Characteristic
Operational Quantity*
Reproducibility
Dose Rate Range
Energy dependence X-/g-rays
Angular dependence
Temperature dependence
Battery voltage dependence
Response time
Position on body**
*) Hp(10): Personal dose equivalent at a depth of 10 mm according to ISO 4037.
**) Position on body: The Hp(10) measurement is only valid for a position on the torso
outside any lead apron or other protection. (To estimate effective dose to a user, one
must use other means to estimate things like the environmental radiation situation and the
effectiveness of protection.)
Document version 1.2
PDM radiology technical data
Measure
Hp(10)
10% or 1 µSv, whatever is greatest
40 µSv/h – 500 mSv/h
+/- 30% within N40 – N120
+/- 5% within +/- 5°
+/- 30% within +/- 50°
+200%/-100% within +/- 90°
+/- 5% within 20-26°C
+/- 25% within 15-35°C
+/- 2% from fully charged until low battery
shutdown
Less than 1s above 100µSv/h, less than 5s
otherwise
On torso outside lead apron