Page 1
Cole-Parmer ® WD-200 Series Paraffin Wax Dispenser Instruction Manual ET0003 Version 1.1...
Page 2
Introduction Thank you for purchasing this Cole-Parmer product. To get the best performance from the equipment, and for your own safety, please read these instructions carefully before use. Before discarding the packaging check that all parts are present and correct. 1 Wax dispenser 2 Instruction manual 3 UK power lead...
General Description The Wax Dispenser is designed for on-demand delivery of molten wax. The novel design of the instrument ensures that it takes up a minimum bench space in the laboratory, whilst still maintaining a large 7.5L capacity tank, for melting up to 6kg of pelletized wax at a time.
Electrical Requirements THIS INSTRUMENT MUST BE GROUNDED Before connection please ensure that the line supply corresponds to the power requirements below: Power: 480W Supply requirements: 110 - 120 V or 220 - 240 V ~ 50/60Hz The unit is provided with three power cables consisting of a UK 3-pin and a “Schuko” 2-pin plug for 230 V installations and a NEMA 5-15 plug for 120 V installations.
Safety Cut-outs The equipment is protected by a miniature circuit breaker and a temperature cut out. The miniature circuit breakers are located on the rear of the equipment. In the event of a fault, push back to reset. If the fault continues, please contact your local service representative or the service department of Antylia Scientific.
Page 6
6. The heater indicator will illuminate to show heater activity. 7. The equipment will warm up to the desired temperature and temperature rise will be shown on the display. 8. The wax dispenser is designed to melt bulk quantities of pelletized wax in as quick a time as possible. When set at 65 C the instrument will take roughly 90 minutes to fully melt 5kg of pelletized wax, with significant quantities of molten wax available for use after as little as 30 minutes.
Maintenance, Servicing and Cleaning WARNING: Before attempting any maintenance, servicing or cleaning, ensure that the unit is cool. WARNING: Do not touch any electrical contacts or open any closure plates. RISK OF ELECTRIC SHOCK. WARNING: Ensure the unit is disconnected from the power supply before attempting any maintenance, servicing or cleaning.
In Case of Contamination WARNING: THE FOLLOWING PROCEDURE IS INTENDED AS A GUIDE. SHOULD SPILLAGE OF A TOXIC OR HAZARDOUS FLUID OCCUR, THEN ADDITIONAL SPECIAL PRECAUTIONS MAY BE NECESSARY. If the equipment has been exposed to contamination, the Responsible Body is responsible for carrying out appropriate decontamination.
Technical Specification General Specification Class 1 cut out Safety functions Thermal cut out 120ºC Tank cover Stainless Steel Tank material Stainless Steel Tank capacity Approx. 7.5 litres Maximum tank capacity External dimensions W181 x D499.5 (inc tap) x H455mm 7.0kg (without wax) Weight 13.0kg (with Max.
Page 11
This product meets the applicable CE Directives that interference will not occur in practise. Where and UKCA Legislation for radio frequency there is a possibility that injury, damage or loss might interference and may be expected not to occur if equipment malfunctions due to radio interfere with, or be affected by, other equipment with similar frequency interference, or for general advise before qualifications.