Take CC mode as an example (other modes operate similarly). When the
measured instrument output voltage is 10V, current is 3A, and load current is
switched between 1A and 2A, the transient test parameters and steps are set as
follows:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10. Press
11. Press any key of
Toggle mode(Toggle)
In toggled transient operation, the load starts at the stored parameters for the
mode. When a trigger is received, the load switches to B value. When another
trigger is received, the load switches to the A level. It stays at the A value until
another trigger is received, at which point it switches to the B value. Here's an
example:
10A
5A
Take CC mode as an example (other modes operate similarly). When the
measured instrument output voltage is 10V, current is 3A, and load current is
switched between 1A and 2A, the transient test parameters and steps are set as
follows:
1.
2.
Press
(Shift)+
confirm.
Press
, and move to PULSE, then press
indicator lamp Trig will be lighted).
Set the rising slope, press
Set the descending slope, press
Set level A, press
Set level B, press
Set the WIDTH, press
Open the transient test function, Press
to confirm.
Then the VFD will display TRAN and Trig.
to turn on the input function, and press
(Trigger) to trigger.
function keys to exit the transient test function. If you need to continue
transient test parameter setting and perform transient test, repeat steps 1 to
10.
TRG
Toggled Transient Operation
Press
(Shift)+
to confirm.
TRAN
On
Off
Press
to select transient operation mode as TOGGLE, then
Copyright © Itech Electronic Co., Ltd.
(Tran) , press
to confirm.
to confirm.
to confirm.
to confirm.
/
/
TRG
(Tran) , move
Functions and Characteristics
and move to ON, press
to confirm (the
to confirm.
and move to on, then press
/
or other composite
key to select ON, press
to
(Shift)+
32