Sb550 20Ma Current Loop Connections - Cardinal SB550 Installation And Technical Manual

Remote display
Table of Contents

Advertisement

SB550 20mA CURRENT LOOP CONNECTIONS

CURRENT Cardinal Indicators
200
PORT 1
P3, 3
P3, 4
PORT 2
P3, 6
P3, 7
205, 210, 210FE, 212/212X, 212G/212GX
PORT 0 (ACTIVE)
P13,2
P13,6
JUMPER
P13,7 to P13,10
PORT 1 (ACTIVE)
P13,2
P13,8
JUMPER
P13,9 to P13,13
225
PORT 0 (ACTIVE)
P20,10
P20,8
PORT 1 (ACTIVE)
P20, 3
P20, 4
JUMPER
P20, 5 to P20, 8
PORT 2 (ACTIVE)
P16, 1
P16, 2
JUMPER
P16, 3 to P16, 9
825
PORT 2 (ACTIVE)
P21, 2
P21, 5
J3 INSTALLED
J7 SHUNT:20mA
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
n/c
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
n/c
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
n/c
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
n/c
SB550
Pin 1, CH1, CL RX+
Pin 2, CH2, CL RX-
n/c
3601-0810-0M Rev C • SB550
7