User Manual I. Disclaimer • The intellectual property rights for this manual belongs to Holland Green Science Corporation (hereinafter referred to as “the Company”). • Products of the company are under the patent protection of China and other countries and regions (inclusive of obtained patents and patents currently under application).
Page 4
User Manual Authorization ("RMA"), which may include biohazard decontamination procedures and other product-specific handling instructions, then, if applicable, Buyer may return the defective Products to the Company with all costs prepaid by Buyer. Replacement parts may be new or refurbished, at the election of the Company. All replaced parts shall become the property of the Company.
User Manual COMPANY HEREIN TO ANY OTHER PARTY OTHER THAN BUYER, INCLUDING BUT NOT LIMITED TO, SUCCESSORS, HEIRS, SUBSIDIARIES OR ANY OTHER THIRD PERSON. THE WARRANTIES IDENTIFIED ABOVE ARE THE SOLE AND EXCLUSIVE WARRANTIES WITH RESPECT TO THIS PRODUCT AND ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED.
User Manual II. Safety IMPORTANT Safety Items to be Observed. Safety Warnings ! DANGER (may cause serious damage to property and or casualties) 1. Please carefully read this User Manual prior to operating the instrument and observe the instructions on safe operation. 2.
User Manual After Sales Support If problems are encountered or technical support is required when installing or using the instrument, please contact serviceusa@hollandgreenscience.com The Company may provide technical assistance and information regarding the instrument or equipment or Service without charge at its sole discretion. Buyer assumes sole responsibility for any reliance on or use of such assistance and information, and the Company makes no warranty thereon.
User Manual III. Aquilon Introduction The Aquilon recirculating chillers feature PID temperature controllers to provide accurate and reliable temperature control and cooling. They are designed for continuous, reliable operation When the liquid is not in use, the liquid should be discharged. Included are alarm functions such as, timing and over-temperature protection.
User Manual IV. Technical Specifications Model 10303001 10303002 10303003 Product Name Aquilon Min Aquilon Aquilon Max Reservoir volume(L) 10.5 ﹣20℃~20℃ Process temperature (℃) ≤30 Ambient Temperature (℃) Refrigeration Ambient relative humidity ≤65 Performance (%RH) Temperature stability (℃) ±0.5℃ Cooling capacity...
User Manual Proper Use The instrument is designed for non-residential use and to be used in conjunction only with accessories recommended within this manual and by the manufacturer. Inspection Unpack the equipment carefully and check for any damage which may have arisen during transport.
User Manual VII. Operating Instructions Preparation and Precautions Prior to Use a. The instrument must be placed on a flat and level surface composed of non-combustible materials. b. 300mm (minimum) of clearance is required around the instrument to ensure the front and back ventilation grilles are well ventilated.
User Manual 7.1 User Interface • TEMP: Displays the measured temperature value. To set the desired temperature, press the SETTINGS key until the TEMP SET value flashes. Press the setting key again to confirm the parameter. • SET: Displays the desired temperature value set manually. •...
User Manual 7.2 Startup Connect the power according to the power outlet. Turn on the instrument at the main switch. The display shows the initialization sequence before the instrument is ready to use. NOTE: Every time you turn on the instrument it will beep intermittently, the icon of visual alarm and the word “end”...
User Manual 7.4 Start / Stop Cooling Operation Before the refrigeration is turned on, please press key to let the liquid circulate in the pipeline in advance. After setting the operating parameters, press button for 4 to 5 seconds to start the cooling process.
User Manual VIII. Troubleshooting Problem Causes Solution The power cord is unplugged Reinsert or not plugged in correctly No power Reset once, if a further trip occurs, Tripped Breaker contact serviceusa@hollandgreenscience.com Maintenance and Cleaning Keep the instrument dry and clean in routine operations. Clean the outer surface with a non-abrasive cleanser and only connect the power supply when the entire instrument surface is dry.