Operating Instructions; Preparation And Precautions Prior To Use - Holland Green Science 10303001 User Manual

Table of Contents

Advertisement

VII.

Operating Instructions

Preparation and Precautions Prior to Use

a. The instrument must be placed on a flat and level surface composed of non-combustible
materials.
b. 300mm (minimum) of clearance is required around the instrument to ensure the front
and back ventilation grilles are well ventilated.
c. This instrument is equipped with a compressor. If the instrument is transported, do NOT
operate immediately. Stand in the correct position or location and wait 1 to 2 days prior
to operating.
d. When filling the liquid tank, ensure liquid does not enter the instrument casing.
e. The liquid must cover the evaporator.
f.
Use an appropriate liquid for desired temperature range
g. When the liquid is water, pay close attention to the temperature set point to prevent Icing
h. Do not mix liquids
DANGER! Do NOT use flammable liquids. Use of such is classified as 'improper use' of
i.
the instrument.
j.
Tighten hose connections securely.
k. During continuous operation, the connection between the bath cover and the circulation
pump may become very cold. When touching these areas ensure suitable PPE is being
worn.
l.
Do not transport the instrument with liquid inside the tank due to the potential of liquid
ingress into the instrument.
m. Use only clean liquids to prevent sediment or solids entering the cooling system and
damaging the instrument.
Version 1.0
User Manual
11

Advertisement

Table of Contents
loading

This manual is also suitable for:

1030300210303003

Table of Contents