Affinity Compact - Instructions for Use - US
The Graph shows measured HIT curves. The X axis shows the
f requency, and the Y axis shows output or gain, depending on
which measurement was done.
Measurement type is printed above the graph together with a
right/lef t indication. In this example the OSPL90 is displayed f or
the lef t ear.
Change the input level using the slider on the right-hand side.
NOTE: f or the industry standard protocols (ANSI and IEC), the
input level is dictated by the standard and cannot be changed.
Scroll graph up/down on the lef t-hand side allows f or scrolling
the graph up or down ensuring that the curve is always visible in
the middle of the screen.
Measurement details: In this table the curve details can always
be viewed. This way the prof essional always has an overview of
what is being measured. Read inf ormation such as Input Level,
Max SPL, Curve Type, Stimulus, and Curve type.
A Curve Comment f or each curve can be typed into the comment
section on the right-hand side.
Select a curve using the curve tag boxes under Curve display
options and write a comment in the comment section.
The comment will then appear in the comment section whenever
the curve is selected.
Curve Display Options are f ound in the lower right-hand corner.
If you have measured more curves of the same type (e.g.
Frequency response curves), they will be listed by their input
level. Tick the ones that are to be displayed on the graph.