Interacoustics Affinity Compact Instructions For Use Manual page 27

Hide thumbs Also See for Affinity Compact:
Table of Contents

Advertisement

5
HF requires an additional license for the AC440. If not purchased, the button is grayed out.
6
MF requires additional license for the AC440. If not purchased, the button is grayed out.
D-0123672-H – 2022/05
Affinity Compact - Instructions for Use - US
Fold an area so that it only shows the label or the buttons of that area.
Unfold an area so that all buttons and labels are visible
Show/hide areas can be f ound by right mouse clicking on one of the
areas. The visibility of the dif ferent areas as well as the space that they
take on the screen is locally saved to the examiner.
List of Defined Protocols allows f or selecting a test protocol f or the
current test session. Right mouse click on a protocol allows the current
examiner to set or deselect a def ault startup protocol.
Please ref er to the Af f inity Compact "Additional Inf ormation" document f or
more inf ormation on protocols and protocol setup.
Temporary Setup allows f or making temporary changes to the selected
protocol. The changes will be valid f or the current session only. Af ter
making the changes and returning to the main screen, the name of the
protocol will be f ollowed by an asterisk (*).
List of historical sessions accesses historical sessions f or comparison
purposes. The audiogram of the selected session, indicated by the orange
background, is shown in colours as def ined by the used symbol set. All
other audiograms that are selected by check marks show on screen in the
colours as indicated by the text colour of the date and time stamp. Note
that this listing can be resized by dragging the double lines up or down.
Go to Current Session brings you back to the current session.
High Frequency shows f requencies on the audiogram (up to 20 kHz f or
the Af f inity Compact
.0
f requency range the selected headset is calibrated f or.
High Frequency Zoom
on the high f requency range.
Single audiogram toggles between viewing the inf ormation of both ears
in a single graph and two separate graphs.
Multi frequencies
6
activate testing with f requencies in between the
standard audiogram points. The f requency resolution can be adjusted in
the AC440 setup.
Synchronize channels locks the two channels together. This f unction
may be used to perf orm synchronous masking.
). However, you will only be able to test in the
activates high f requency testing and zooms in
5