7.2 Replacing the LC163 Ion-exchange
Cartridge Pack
The ion-exchange cartridge pack should be replaced in the following
circumstances:
 The water purity alarm monitor indicates that the pack
requires changing.
 If the system is being re-commissioned or sanitized after an
extended period in which it was not used.
 When indicated by the change reminder.
Step 1 - Switch Unit Off
1. SWITCH the PURELAB Option-Q off at the
power switch at the top left hand side of the unit.
2. OPEN the dispense tap to relieve any residual
pressure from the system.
WARNING!
Step 2 - Remove Ion-exchange Cartridge
1. OPEN the front door.
2. PUSH on cartridge top cap.
3. LIFT up cartridge.
4. SLIDE out cartridge.
5. DISCARD used ion-exchange cartridge.
Note:
Step 3 - Replace Ion-exchange Cartridge
1. REMOVE a new cartridge pack
packaging.
2. REMOVE the sealing plugs from the inlet and
outlet ports.
3. WET 'O' rings and SLIDE new cartridge into
position.
4. POSITION cartridge onto spigots, PUSH into
unit.
5. ENSURE guide has dropped down past retainer.
6. CLOSE front door.
7. RESET ion-exchange pack timer. (See Section
5.4 - Setting up Display and Replacement Timer).
8. PRESS the process button to start the unit.
9. CHECK the unit for leaks.
PURELAB Option-Q 7/15 - US Version 3 02/16
ENSURE THE UNIT IS ISOLATED BEFORE
REMOVING
THE
CARTRIDGE.
The consumable is non-hazardous.
Dispose
of
as
ordinary
observing
all
local
regulations.