Delete tare
PRE-TARE
function
Delete PRE-TARE
value
EMS-BA-e-1112
Weigh the material, the net weight will be indicated.
The taring process can be repeated any number of times,
e.g. when adding several components for a mixture
(adding). The limit is reached when the whole weighing
range is exhausted.
The weight of the weighing container will be displayed as a
minus number after removing the weighing container.
The tare weight is saved until it is deleted.
Unload the balance and press the TARE button, the zero
display appears.
Using this function the weight of a tare vessel is stored. Even
after turning off/on the weighing balance will continue working
with the saved tare value.
In weighing mode put tare vessel on the weighing plate
Press repeatedly the MODE key until „PtArE" flashing
appears.
Use SET key to store the current weight on the weighing
plate as a PRE-TARE value.
Remove all loads from the balance, press TARE and press
repeatedly the MODE key until „PtArE" flashing appears.
Use the SET key to confirm. The PRE-TARE value is
deleted, the zero display appears.