Establishing the daily operating reference density levels for
mammography application
This procedure enables you to establish the base line values for:
• Base + Fog
• Low density
• Mid density
• High density
To establish the daily operating levels, proceed as follows:
1. Press the Key-operator key to enter the Key-operator mode.
2. Press the down key seven times, followed by the confirm key to select
'Quality Control'.
5 Restore config.
6 Calibration
7 Service Actions
8 Quality Control
9 Installation
A confirmation screen appears:
3. Do one of the following:
• Press the Confirm key (YES) to start the QC procedure and proceed
with step 4.
• Press the Escape key (NO) to quit.
4. Define the tray for printing.
The 'Select tray' screen appears:
5. Press the Up/Down arrow keys to select the desired tray, followed by the
Confirm key.
The following screens appear successively:
Drystar AXYS | Advanced Operation (key-operator mode) | 111
WARNING:
The densitometer of the printer is calibrated at installation.
Authorized service personnel should recalibrate the
densitometer annually or after major service or repair.
Visible
Reachable with Up/Down arrow keys
2852E EN 20210601 1655