1 About MIWI350
1.4.6 Temperature Measurement
• Highly stable, solid state temperature sensor (Thermistor)
• Maximum error over the full temperature range -40° to 70°C (-40 °F to +158 °F) including
linearity, repeatability and hysteresis:
– Temperature: ±0.1% of reading
• Maximum error at reference conditions:
– Temperature: ±0.1% of reading
1.4.7 Pressure Measurement
Ambient temperature range of Pressure Transducer:
• Maximum error over the full temperature range -40° to 70°C (-40 °F to +158 °F) including
linearity, repeatability and hysteresis:
– Pressure: ±0.1% of full scale
• Maximum error at reference conditions:
– Pressure: ±0.04% of full scale
• Long term drift:
– Pressure: ±0.4% of full scale over 10 years
Maximum pressures per pressure transducer type and range: 6 to 1500psi
Note:
For more specification details, please refer
Type
Gauge
Gauge
Gauge
Gauge
Gauge
Gauge
Gauge
Gauge
Absolute
Absolute
Absolute
Absolute
Absolute
Absolute
Pressure Expansion (PE) Board for Pressure Transducer, P3:
g
EC350
Datasheet.
Pmax
6
30
60
100
300
600
1000
1500
30
100
300
600
1000
1500
/30-1500psi
a
Units
psi
g
psi
g
psi
g
psi
g
psi
g
psi
g
psi
g
psi
g
psi
a
psi
a
psi
a
psi
a
psi
a
psi
a
|6