Do not obstruct the mains power switch or
WARNING
position the equipment where the connection to
the mains line can be accidentally
disconnected.
Equipment is not suitable for use in the
WARNING
presence of a Flammable Anesthetic Mixture
with Air, Oxygen, or Nitrous Oxide.
The Optomed Polaris has no special protection
WARNING
against harmful ingress of water or other liquids
(classified IPX0). To avoid damage to the
instrument and cause a safety hazard, the
cleaning solutions, including water, should not
be directly applied to the device. Using a
dampened cloth (without dripping), is a good
method to clean the exterior surface of the
enclosure.
The patient cannot touch any electrical device
WARNING
that is not powered by Optomed Polaris with
any part of his or her body while being
examined. In addition, the Optomed Polaris
operator must not attempt to touch the patient
and any electrical device that is not powered by
Optomed Polaris at the same time while
examining the patient. Failure to do so could
result in electrical shock to the patient and/or
operator.
Do not connect the instrument with anything
WARNING
other than specified. Otherwise, it may result in
fire or electric shock. For details of purchasing
accessories, please contact a Optomed
representative or distributor.
Be sure to hold the bottom of the base when
CAUTION
Optomed Polaris is moved.
Non-Mydriatic Fundus Camera (Optomed Polaris) - User Manual