Page 3
Optomed Aurora User manual for Optomed Aurora ophthalmic camera Manufacturer: Optomed Oy (Ltd) Yrttipellontie 1, FI-90230 Oulu Finland Tel: +358 20 741 3380 info@optomed.com English 60000888 R 6.0 www.optomed.com 2019 07 01...
Controls and display icons Getting started Part list and unpacking Preparations Features Quick menu Camera settings Optics settings for Aurora Retinal Module Optics settings for Aurora Anterior Module Imaging with Aurora Retinal Module Anterior Module After imaging Important information Environmental conditions...
The user manual and other documentation enclosed with the Optomed Aurora should be kept accessible to users at all times to ensure that the information required for use of the device is readily available.
The eye cup which is in physical contact with the patient, is electrically isolated and protected against electric shock. Class II power supply For indoor use only Power supply designed primarily for indoor use only according to IEC 60417 Safety | Optomed Aurora...
Page 8
Serial number The first four digits of the serial number indicates the week (digits 1-2) and year (digits 3-4) of the manufacture. Global trade item number Reference number GS1 Unique Device Identification Keep away from water Optomed Aurora | Safety...
Page 9
Fragile Handle with care. Do not use hand hooks Keep out of direct sunlight Number of units packed Temperature limitations Humidity limitations Atmospheric pressure limitations Safety | Optomed Aurora...
1.3 Contraindication Optomed Aurora is classified as Group 2 based on standard ISO 15004-2:2007. The daily usage time and maximum allowed number of pulses presented below is calculated on the basis of optical classification results.
1.4 Precautions WARNING! Optomed Aurora is not suitable for use in explosion risk areas or in the presence of flammable anesthetics. WARNING! Place the PC and Aurora Charging Station outside the patient environment (at a distance of at least 1.5 meters from the patient).
Page 12
PC. If the PC is connected to a customer network, make sure the appropriate safety measures are applied, such as Antivirus and Firewall protection. CAUTION! If there are breaks in device covers or other visual defects, contact Optomed Customer Service or an Optomed-certified service facility. CAUTION! Repair and maintenance work may only be performed by authorized specialists.
Aurora Camera has a WLAN module which enables wireless data transfer to a PC. Captured images and recorded videos can also be transferred to a PC via USB connection when the camera is placed on Aurora Charging Station. In addition, Aurora Charging Station is a dual charger used for charging the battery inside the camera and spare battery.
Page 14
Patient Battery level Fixation target WLAN indicator Brightness level Diopter value Power button Back button Optoroller Battery charging indicator LED Dual action shutter Eject button Optomed Aurora...
Battery level, WLAN, patient, fixation target, brightness level and diopter value indicators are shown on the camera display. Aurora Camera is powered on and off by pressing the power button. Opening the camera menu is done by pressing Optoroller, which is also used for navigating in the menu together with back button.
Cleaning Cloth User Manual and other material Open the sales packaging. Remove Optomed Aurora and accessories from the sales package and check that all parts are present and undamaged. Packaging materials should be retained for future relocation or repair. 3.2 Preparations Place Charging Station (4) next to the PC.
Page 17
Place the eye cup (7) on the optics ring of Aurora Retinal Module (2). Install the battery (8) in Aurora Camera (1). Ensure that the camera is powered off when placing the battery. Open the battery cover by sliding the battery cover release knob.
Page 18
PC. If Aurora Camera is not used for more than two minutes, the device will go into power save mode. It is powered on from power save mode by using any control button. Aurora Camera will power off if not used for more than 10 minutes.
The autofocus range is from -15 to +10 diopters. BRIGHTNESS Imaging light brightness for Aurora Retinal Module can be adjusted from 0 to 10 and the default value is 5. It is also possible to reduce the brightness setting below 1, to 0.8, 0.6, 0.4, and 0.2, if needed.
Page 20
Optoroller. It is also possible to delete complete studies or single images. When imaging with Optomed Aurora Retinal Module, there is a possibility to select the ‘ Four image sequence ’ item by pressing Optoroller. By selecting four image sequence, a new patient study is created and the camera will guide to take both macula and disc centered images from patient’s right and left eyes.
WIRELESS Optomed Aurora has a WLAN module that enables wireless connection to a PC. To change wireless settings, press Optoroller. Appendix A provides instructions for wireless settings.
Page 22
’ is selected, the camera will guide to take both macula and disc centered images from patient’s right and left eyes. If ‘ No query’ is selected, the camera will go directly to live view after powering on or when removed from Aurora Charging Station. Optomed Aurora | Features...
• Autofocus: image is captured after focusing when the dual action shutter is kept pressed half way down CAPTURE MODE There are two capture modes in Optomed Aurora: still imaging and video recording. To choose capture mode, use Optoroller. ADDITIONAL IMAGE In addition to color image, there are two alternative imaging modes in Optomed Aurora: Red-free and Low-red.
Page 24
Much reductions in image CENTERING Centering allows adjustment of the imaging area to correspond with the view area. First select vertical or horizontal adjusting by using Optoroller. Move the image to the middle of the screen by turning Optoroller. Optomed Aurora | Features...
• Autofocus: image is captured after focusing when the dual action shutter is kept pressed half way down CAPTURE MODE There are two capture modes in Optomed Aurora: still imaging and video recording. To choose capture mode, use Optoroller. Features |...
3. Respectively, brightness can be increased up to 10, if images are too dark. • When imaging with Aurora Anterior Module, first select the imaging light to be white or blue and then set the brightness. The default value is 5, which is suitable for most imaging situations.
5.1 Retinal Module When imaging with Aurora Retinal Module, the examination room should be as dim as possible. It is recommended that both the patient and the examiner are seated during the examination. It is also possible to perform the examination when the patient is lying down.
5.2 Anterior Module It is recommended that both the patient and the examiner are seated during imaging with Aurora Anterior Module. Place the eye cup of Anterior Module around the patient’s eye. Stabilize the camera by supporting the optics module on your thumb and place your fingers on the patient’s forehead.
The camera will continue wireless transfer when removed from Aurora Charging Station. Images and videos can also be transferred to PC from the memory card when the camera is placed on Aurora Charging Station. Aurora Camera will display a ‘Connected’ message when the device is connected to a PC via USB.
Protection against harmful ingress of water or particulate matter is defined as IPX0 (Equipment not protected against the ingress of water). 6.2 Error messages Optomed Aurora will display error messages to indicate error situations. An error message is always displayed with an explanatory message providing information about possible actions. ERROR MESSAGE...
Page 31
‘Erase image memory’ from Camera settings. If the device behaves abnormally, Aurora Camera can be forced to shut down by pressing the power and back buttons simultaneously for seven seconds or longer.
• Nail polish remover Use of ammonia-based cleaners on the liquid crystal display (LCD) may cause damage to the display. The Optomed Aurora is a precision optics instrument that should be handled with care. Please note the following cleaning instructions: •...
Aurora Camera has a rechargeable Li-ion battery that is charged when the camera is placed on Aurora Charging Station, which is connected to the mains. Aurora Charging Station can also be used as an external battery charger for the spare battery included in sales case. The battery pack is specially designed and manufactured for this device.
For further information on disposal of this product, please contact your Optomed representative. This will guarantee that no hazardous substances enter the environment and that valuable raw materials are recycled.
Page 35
Usage time: Approximately 2 h Device life time: Approximately five years OPTICS FOR RETINAL IMAGING Model: Optomed Aurora Retinal Module Dimensions: 69 (w) x 74 (h) x 160 (d) mm Weight: 324 g Illumination: Infrared LED for aiming. Visible white LED for imaging, 10 illumination brightness levels.
Page 36
2368 x 1776 pix (total 4.2 Mpix, effective area 3.38 Mpix) Spectral output at working distance: White and Infrared LED when operated in continuous mode OPTICS FOR ANTERIOR IMAGING Model: Optomed Aurora Anterior Module Dimensions: 71 (w) x 80 (h) x 78 (d) mm Weight: 105 g...
USB cable: Type: A to mini B Length: 1.5 m MEMORY STICK Memory stick including Optomed Client PC Software and Instructions for use. CLEANING CLOTH Cleaning cloth for cleaning lens and display. INTELLECTUAL PROPERTY RIGHT INFORMATION Windows 7, Windows 8.1 and Windows 10 are trademarks of Microsoft Corporation.
Page 38
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Optomed Aurora including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Page 39
8 ф angles; at 8 ф angles; typical commercial or hospital environment. and voltage If the user of the Optomed Aurora requires 0% 1 cycle; 0% 1 cycle; variations on continued operation during power mains 70% UT for 25...
Page 40
If the measured field strength in the location in which the Optomed Aurora is used exceeds the applicable RF compliance level above, Optomed Aurora should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Optomed Aurora.
Optomed Aurora maintains basic safety and performance when used in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of Optomed Aurora can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and Optomed Aurora as recommended below, according to the maximum output power of the communications equipment.
Optomed Aurora is covered by a limited warranty granted by Optomed Oy. SUBMITTING A CLAIM Any claim under this warranty must be submitted in writing before the end of warranty period to Optomed Oy. The claim must include a written description of the failure to the device.
A. Appendix - WLAN A.1 Wireless settings Aurora Camera has a WLAN module that enables wireless connection to a PC or other client device. In order to change wireless settings, open the wireless menu from camera settings by using Optoroller...
A.2 Optomed Client PC software The PC software Optomed Client enables WLAN connectivity from PC or other client device to Aurora Camera. With Optomed Client software you can wirelessly connect PC or other client device to Aurora Camera and define the image destination folder into which images will be transferred.