Unsuccessful Calibration

A-61408
September 1995
If calibration fails, perform the following steps:
1. Verify that you are using a clean, matte (non-glossy), blank
sheet of paper to calibrate the scanner.
2. Make sure that length of calibration target is
3. Make sure no paper is in the document path. Refer to the
section entitled, Clearing the Document Path .
4. Clean the imaging guides. Refer to the section entitled,
Cleaning the Imaging Guides .
5. Calibrate the scanner again. If this calibration fails, change
the lamps. Refer to the section entitled Changing the Lamps .
Be sure to calibrate the scanner after changing lamps.
If calibration still fails, contact your local service representative.
12".