GE Limitamp CR194 Installation Instructions Manual page 22

Two-high vacuum control
Hide thumbs Also See for Limitamp CR194:
Table of Contents

Advertisement

CR194 Two-High
Vacuum Limitamp@ Control
CONTROL
POWER INTERLOCK
(CPI)
Vacuum Interrupter Integrity Test
WARNING:
Confirm
that equipment
is de-ener-
gized before
performing
this check.
Check
the CPI adjustments
as shown in Figure
42 and
adjust if required.
1 \
a
;
HANDLE
I-Em4
+f
LATCH
RELEASE
9
c-
--
J
NOTE:
1. SET 0.12~INCH GAP (AS SHOWN) WITH PUSHER ROD
AND HANDLE LATCH RELEASE EXTENDED
(OUT
POSITION). SEE 0.60-INCH REFERENCE.
2. THEN PUSH IN THE HANDLE LATCH RELEASE AND
CHECK THE CPI CONTACT GAP
3. BOTH CONTACTS MUST BE OPEN AND HAVE A
CONTACT GAP OF AT LEAST 5/64 INCH
Figure 42. Checking for proper operation of pusher rod
and control power interlock (CPI)
TEST POWER INTERLOCK
(TPI)
Check
the TPI adjustments
as shown in Figure
43 and
adjust if required.
TPI
INTERLOCK
\
OPERATING
ROD
REF: FIG. 39
FOR
SIDE VIEW
DJUST INTERLOCK DEPRESSOR BRACKET(S) BY BENDING TO DEPRESI
INTERLOCK PLUNGERS FULLY WITH HANDLE IN "OFF" POSITION
Figure 43. Checking test power interlock (TPI) adjustment
WARNING:
The vacuum
interrupter
integrity
test
should
be performed
before
the high-voltage
contactor
is energized
for the first time and each
time it is returned
to service
after maintenance,
adjustment
or repair
(See GEH-5306).
Otherwise
the vacuum
interrupter
integrity
test should
be
performed
annually.
Failure
to perform
this test
may cause
serious
injury
or death.
WARNING:
All barriers
must be in place
and the
isolation
switch handle
must be in the "OFF"
(down)
position
(i.e. the handle
latch
release
pops
back out)
before
attempting
to perform
the
vacuum
interrupter
integrity
test.
WARNING:
This high-voltage
test is potentially
hazardous,
so extreme
caution
should
be exer-
cised.
First open
the controller
isolator
switch,
then completely
isolate
the contactor
by discon-
necting
the contactor
line side cables
and load
side bolts and sliding
the contactor
to the "inter-
mediate"
position
using
the safety stop.
WARNING:
X-Ray emission
may be produced
if an
abnormally
high voltage
is applied
across
the open
contacts
of a vacuum
interrupter.
Do not apply an
RMS voltage
across
the contactor
that is higher
than 20.0 kV RMS.
NOTE:
Before performing this vacuum integrity test, confirm
that both the armature gap setting and contact wear adjust-
ment are proper (See GEH-5306,
Sections 5&6).
GENERAL
This test determines
the internal
dielectric
condition
and vacuum
integrity
of the vacuum
interrupters.
Prior
to performing
this test, the outside
surfaces
of
the vacuum
interrupters
should
be wiped clean
of any
contaminants
with a non-linting
cloth
or industrial
type wiper.
During
this test, each interrupter
should
be checked
separately.
High potential
test instruments
can be purchased
to
perform
the vacuum
interrupter
integrity
test. The fol-
lowing
is a recommended
test instrument:
Hipotronics
Model
7BT60A
Use of a DC Hipot
is not recommended
because
re-
sults may indicate
a problem
with a good interrupter.
If you wish to use a DC Hipot,
set for 28 kV, but if an
interrupter
fails, confirm
the failed
interrupter
using
above AC Hipot.
NOTE:
If the controller fails to meet any of the above Me-
chanical Operation Checks, DO NOT ENERGIZE THE
CONTROLLER,
Contact your nearest GE sales office.
24

Advertisement

Table of Contents
loading

Table of Contents