Power Requirements; Connections - Boonton SGX1000 Series Instruction Manual

Rf signal generator
Hide thumbs Also See for SGX1000 Series:
Table of Contents

Advertisement

the bottom of the instrument. Rack mounting instructions are provided with the optional rack mount
kit.

2.2 Power Requirements

The SGX1000 series is equipped with a switching power supply that provides automatic operation from a
90 to 264 VAC, 47 to 63 Hz single-phase, AC power source. Maximum power consumption is 45 VA.
Connect the power cord supplied with the instrument to the power receptacle on the rear panel. See
Figure 3-2.
Caution
Before connecting the instrument to the power source, make certain that a 1.0-ampere
time delay fuse (type T) is installed in the fuse holder on the rear panel.
Before removing the instrument cover for any reason, position the input module power
switch to off (0 = OFF; 1 = ON) and disconnect the power cord.

2.3 Connections

USB
Two standard USB 2.0 Type A receptacles are located on the front and rear panels of the
instrument and accept standard USB keyboards, mice, and flash drives.
RF OUT
The RF output is available from a Type-N connector located on the front, or optionally
on the rear panel of the instrument. A blue LED indicator is illuminated when the RF
output is active.
Ref In
BNC connector for 10 or 100 MHz reference input.
Ref Out
BNC connector for 100 MHz reference output.
Trig In
BNC input for connecting an external trigger signal to the signal generator.
Multi-I/O
BNC connector reserved for future use.
LAN
LAN/Ethernet connector for remote control. Allows DHCP or fixed (IP / Subnet) setting
mode. In most cases, it will be necessary to configure the interface using the
System→I/O Config menus.
Caution
The supplied Ethernet cable fitted with a ferrite core must be used for RF interference
suppression
GPIB
Optional 24-pin IEEE-488 General Purpose Interface Bus (GPIB) connector for remote
control of the signal generator. GPIB parameters can be configured through the menu.
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

98409500a