Siemens SITRANS FS230 Compact Operating Instructions page 16

Ultrasonic flowmeters
Table of Contents

Advertisement

4.4
HI Precision mount
Single enclosure Reflect mount wiring
Connect sensor cables to the transmitter as follows:
1. Fill connector end with Super Lube prior to connecting.
2. Apply Super Lube to the internal threads at the large end of the thread connector.
3. Observing the upstream and downstream orientation
make connection snug.
4. Connect downstream sensor cable
Wall mounted transmitter
Upstream sensor cable to internal DSL port
(1A) Path 1
Figure 4-3
Hi Precision Reflect mount single enclosure wiring
5. Tighten all cable glands to obtain optimum sealing.
SITRANS FS230
A5E38755607-AA, 02/2017
, connect upstream sensor cable
to the transmitter port (1B) and make connection snug.
Flow direction
Downstream sensor cable to transmitter port (1B) Path 1
Hi Precision mounted sensors
to transmitter port (1A) and
15