1.7
Wire detector
To install the detector wiring:
Figure 1-5: Detector with Cover Removed
A. Housing
B. Internal earth connection
C. Terminal board
D. Inlet conduit
E. Detector holding plate
F. Earth terminal
G. Connection to Field Communicator
Procedure
1. Release the back cover secure bolt and open the detector back cover.
2. Remove the protective plug mounted on the detector conduit/cable
entry inlet.
3. Pull the wires through the detector inlet.
4. Use a ¾-in.-14 national pipe thread (NPT) or M25 x 1.5 explosion-
proof conduit connection/cable gland to assemble the cable/
explosion-proof conduit to the detector.