Products Solutions Services KA01510D/06/EN/01.21 71542316 2022-01-11 Brief Operating Instructions Proline 400 Transmitter with ultrasonic transit time sensor HART These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Brief Operating Instructions part 2 of 2: Transmitter Contain information about the transmitter.
Page 2
Proline 400 HART Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
Page 3
These Brief Operating Instructions are Brief Operating Instructions Part 2: Transmitter. The "Brief Operating Instructions Part 1: Sensor" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer •...
Result of a step Visual inspection 1.1.3 Electrical symbols Symbol Meaning Symbol Meaning Direct current Alternating current Direct current and alternating current Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system. Endress+Hauser...
Page 6
Allen key Open-ended wrench 1.1.6 Symbols in graphics Symbol Meaning Symbol Meaning 1, 2, 3,... Item numbers … Series of steps A, B, C, ... Views A-A, B-B, C-C, ... Sections Hazardous area Safe area (non-hazardous area) Flow direction Endress+Hauser...
If the temperature of the media or electronics unit is high or low, this may cause the surfaces of the device to become hot or cold. This poses a risk of burns or frostbite! ‣ In the case of hot or cold medium temperatures, install appropriate protection against contact. Endress+Hauser...
(Statutory Instruments). These are listed in the UKCA Declaration of Conformity along with the designated standards. By selecting the order option for UKCA marking, Endress+Hauser confirms a successful evaluation and testing of the device by affixing the UKCA mark. Contact address Endress+Hauser UK: Endress+Hauser Ltd.
For detailed information on the product description, see the Operating Instructions for the device → 3 Mounting For detailed information about mounting the sensor, see the Sensor Brief Operating Instructions → 3 Endress+Hauser...
2.5 Nm (1.8 lbf ft) – plastic housing 1 Nm (0.7 lbf ft). Special mounting instructions 4.2.1 Display guard ‣ To ensure that the display guard can be easily opened, maintain the following minimum head clearance: 350 mm (13.8 in) Endress+Hauser...
• Is the fixing screw firmly tightened? • Is the connection compartment cover screwed on tightly? • Is the securing clamp firmly tightened? Turning the display module: • Is the connection compartment cover screwed on tightly? • Is the securing clamp firmly tightened? Endress+Hauser...
Current output 0/4 to 20 mA Standard installation cable is sufficient. Current output 4 to 20 mA HART A shielded cable is recommended. Observe grounding concept of the plant. Pulse/frequency/switch output Standard installation cable is sufficient. Status input Standard installation cable is sufficient. Endress+Hauser...
Page 13
Frequency range "Power supply" DC 24 V ±25% – Option L 1 (L+/L), 2 (L-/N) AC 24 V ±25% 50/60 Hz, ±4 Hz (wide range power unit) AC 100 to 240 V –15 to +10% 50/60 Hz, ±4 Hz Endress+Hauser...
Page 14
Remove dummy plug if present. If the measuring device is supplied without cable glands: Provide suitable cable gland for corresponding connecting cable. If the measuring device is supplied with cable glands: Observe requirements for connecting cables → 12. Endress+Hauser...
Connect the sensor and transmitter to the same potential equalization. ‣ Only connect the sensor to a transmitter with the same serial number. The following sequence of steps is recommended when connecting: Mount the sensor and transmitter. Connect the sensor cable. Connect the transmitter. Endress+Hauser...
Page 16
Connect sensor cable to channel 1 downstream. For a two-path measurement: proceed as per steps 3+4 Connect sensor cable to channel 2 upstream. Connect sensor cable to channel 2 downstream. 10. Tighten the cable gland(s). This concludes the process for connecting the sensor cable(s). Endress+Hauser...
Page 17
Proline 400 HART Electrical connection WARNING Housing degree of protection may be voided due to insufficient sealing of the housing. ‣ Screw in the screw without using any lubricant. Reverse the removal procedure to reassemble the transmitter. Endress+Hauser...
Page 18
Connecting the supply voltage and 0-20 mA/4-20 mA HART with additional outputs and inputs Loosen the 4 fixing screws on the housing cover. Open the housing cover. Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable entry. Endress+Hauser...
Ground the cable shield at one end. The cable shield must be grounded at both ends to comply with EMC requirements; observe cable specifications Connection for HART operating devices Resistor for HART communication (≥ 250 Ω): observe maximum load Analog display unit: observe maximum load Transmitter Endress+Hauser...
Page 20
Connection example for pulse/frequency output (passive) Automation system with pulse/frequency input (e.g. PLC) Power supply Transmitter: Observe input values Switch output A0028760 5 Connection example for switch output (passive) Automation system with switch input (e.g. PLC) Power supply Transmitter: Observe input values Endress+Hauser...
Tighten all housing screws and screw covers. Firmly tighten the cable glands. To ensure that moisture does not enter the cable entry, route the cable so that it loops down before the cable entry ("water trap"). A0029278 Endress+Hauser...
Does the supply voltage match the specifications on the transmitter nameplate ? Is the terminal assignment correct → 13? If supply voltage is present, do values appear on the display module? Are all housing covers installed and the screws tightened with the correct tightening torque? Endress+Hauser...
Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Communicator 475 Field Xpert SFX350 or SFX370 Field Xpert SMT70 Mobile handheld terminal Control system (e.g. PLC) Endress+Hauser...
The individual parts of the operating menu are assigned to certain user roles (operator, maintenance etc.). Each user role contains typical tasks within the device lifecycle. For detailed information on the operating philosophy, see the Operating Instructions for the device. Endress+Hauser...
The Web browser setting Use proxy server for LAN must be disabled. browser JavaScript JavaScript must be enabled. If JavaScript cannot be enabled: enter http://192.168.1.212/basic.html in the address line of the Web browser. A fully functional but simplified version of the operating menu structure starts in the Web browser. Endress+Hauser...
Page 26
Configure the properties of the Internet protocol (TCP/IP) as defined in the table: IP address 192.168.1.XXX; for XXX all numerical sequences except: 0, 212 and 255 → e.g. 192.168.1.213 Subnet mask 255.255.255.0 Default gateway 192.168.1.212 or leave cells empty Endress+Hauser...
Page 27
(e.g. tag name) as it is displayed as the WLAN network. Disconnecting ‣ After configuring the device: Terminate the WLAN connection between the operating unit and measuring device. Starting the Web browser Start the Web browser on the computer. Endress+Hauser...
Page 28
• Access to the operating menu from the measuring device • The structure of the operating menu is the same as for the local display Menu For detailed information on the structure of the operating menu, see the Operating Instructions for the measuring device Endress+Hauser...
Page 29
The Web server of the measuring device can be switched on and off as required using the Web server functionality parameter. Navigation "Expert" menu → Communication → Web server Parameter overview with brief description Parameter Description Selection Web server functionality Switch the Web server on and off. • Off • On Endress+Hauser...
For detailed information on system integration, see the Operating Instructions for the device → 3 • Overview of device description files: • Current version data for the device • Operating tools • Measured variables via HART protocol • Burst mode functionality in accordance with HART 7 Specification Endress+Hauser...
If nothing appears on the local display or a diagnostic message is displayed, refer to the Operating Instructions for the device → 3 Setting the operating language Factory setting: English or ordered local language Endress+Hauser...
Parameters → 3 Submenu Configuration System Display, diagnostic settings, administration Sensor Measured values, system units, process parameter, sensor adjustment Measuring point Configuration of the measuring point Installation status Configuration of the installation status Input Status input Output Current output, pulse/frequency/switch output Endress+Hauser...
Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. The message about remedial measures can be called up from the diagnostic message, and contains important information on the fault. Endress+Hauser...
Page 34
The Diagnostic list submenu opens. Select the desired diagnostic event with or and press . The message about the remedial measures opens. Press + simultaneously. The message about the remedial measures closes. Endress+Hauser...