From step 11
13. Unplug connector 2a from connector 9a. See diagram 3.
14. Measure voltage across pin 1 and pin 4 of connector 9a.
See figure 9.
— If the multimeter indicates 0VDC, then replace the
power module (9) and retest the system.
— If the multimeter indicates greater than 0VDC, then
plug connector 2a back into connector 9a and go to the
next step.
15. Unplug connector 2b from connector 1b. See diagram
2 or 3.
NOTE: If the chair is equipped with a power positioning
system, refer to the troubleshooting guide for that particular
power positioning system.
16. Measure voltage across pin 1 and pin 4 of connector 2b.
See figure 10.
— If the multimeter indicates 0VDC, then replace the bus
cable extension (2) and retest the system.
— If the multimeter indicates greater than 0VDC, then
replace the NE+ joystick module (1) and retest the
system.
8
www.pridemobility.com
Figure 9. Connector 9a
Figure 10. Connector 2b
Quantum 614 with NE+ Electronics Technical Troubleshooting Guide