Appendix A: Zero Set, Zero Drift, and Measurement Noise

82
The following specifications in
to the U2000 Series USB power sensors with the serial
prefix as shown below:
Model
Serial Prefix
U2000A
Serial Prefix MY470/SG470 and below
U2001A
Serial Prefix MY471/SG471 and below
U2002A
Serial Prefix MY472/SG472 and below
U2004A
Serial Prefix MY474/SG474 and below
Table 5-1 Zero Set, Zero Drift, and Measurement Noise for Average only
Mode
1
Range
–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
1 Condition: (i) 0 °C to 55 °C and (ii) 40 °C, 95% relative humidity.
2 Within one hour after zero set, at a constant temperature, after 24-hour warm-up
of the power sensor.
3 The number of averages at 16 for Normal speed, measured over one minute
interval and two standard deviations.
Table 5- 1
Zero Set
Zero Drift
±651 pW
996 pW
±1.13 nW
400 pW
±12.8 nW
6.01 nW
±445 nW
155 nW
±4.26 µW
3.20 µW
±6.84 µW
3.39 µW
Agilent U2000 Operating and Service Guide
are only applicable
2
Measurement
3
Noise
1.91 nW
2.24 nW
40.8 nW
1.63 µW
861 nW
19.5 µW