Switching Point
The U2000 Series USB power sensors have two measurement paths: a low
power path and a high power path as shown in
Table 3-3 Switching point
Sensor
U2000A, U2001A, U2002A, U2004A
U2000H, U2001H, U2002H
U2000B, U2001B
The power sensor automatically selects the proper power level path. To
avoid unnecessary switching when the power level is close to the
switching point, Switching Point Hysteresis is added.
U2000 Series Operating and Service Guide
Low Power Path
–60 dBm to –7 dBm
–50 dBm to +3 dBm
–30 dBm to +23 dBm
Offset at Switch Point
Switching Point Hysteresis
Specifications and Characteristics
Table 3-
3.
High Power Path
–7 dBm to +20 dBm
+3 dBm to +30 dBm
+23 dBm to +44 dBm
Error
≤±0.5% (≤±0.02 dB) typical
0.5 dBm typical
3
Switching Point
–7 dBm
+3 dBm
+23 dBm
51