Vapormed Mighty Plus MEDIC Instructions For Use Manual page 40

Hide thumbs Also See for Mighty Plus MEDIC:
Table of Contents

Advertisement

INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
Immunity
test
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagne-
tic propagation is affected by absorption and reflection from
structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio
a)
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the
MIGHTY + MEDIC Vaporizer is used exceeds the applicable RF compliance
level above, the MIGHTY + MEDIC Vaporizer should be observed to verify
normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the
MIGHTY + MEDIC Vaporizer.
Over the frequency range 150 kHz to 80 MHz, field strengths should be
b)
less than 3 V/m.
IEC 60601
test level
A-6
Compliance
level
Electromagnetic
environment –
guidance
Where P is the
maximum output
power rating of the
transmitter in watts
[W] according to the
transmitter
manufacturer and d
is the recommended
separation distance
in meters [m].
Field strengths from
fixed RF
transmitters, as
determined by an
electromagnetic site
survey
, should be
a)
less than the com-
pliance level in each
frequency range
.
b)

Advertisement

Table of Contents
loading

Related Products for Vapormed Mighty Plus MEDIC

This manual is also suitable for:

Mighty+ medic

Table of Contents