4 USING THE ELCOMETER 138 BRESLE SALT METER (continued)
Correct procedure for closing meter after removal of the sensor:
Ensure that the sensor tongue is outside the meter case. If the tongue is inserted between the case and
the connector of the meter, it may damage the connector.
Sensor Tongue
The sensor is a consumable part. If it becomes damaged or its performance deteriorates, it will need to
be replaced.
Description
Replacement Sensor
4.4 ELECTRODE SURFACE TREATMENT PROCEDURE
When using the sensor for the first time or again after several weeks of disuse, perform the electrode
surface treatment procedure as follows:
1
Put some drops of the conditioning solution into the measurement cell.
2
Leave for approximately 10 minutes.
3
Clean the measurement cell with running water.
4
Wash the measurement cell with standard solution.
5
Perform the calibration procedure, see Section 4.5 on page en-8.
en-7
Meter
Connector
Sensor
ü
Meter
Sensor
û
Meter
Part Number
T13830628