T-BERD
/MTS 5800 Modular Test Set
Real Time Testing with Expert OTDR
This quick card describes how to connect to a fiber under test, configure Expert OTDR settings, run
Real Time tests, and analyze results with a VIAVI T-BERD/MTS 5800 and 4100-series OTDR module.
Equipment Requirements:
• T-BERD/MTS-5800 with
Release V21.04 or greater
E4100 Series OTDR Module
• Fiber optic cleaning and inspection tools
• Launch Cable with connectors matching the
OTDR port and Fiber Under Test (a minimum
20-meter Fiber optic patch cable or leash is
recommended)
• Optical Coupler to connect Launch Cable to
Fiber Under Test
The following information is required to complete the test:
• Type of Fiber (Multimode or Single Mode)
• Type of Connectors (SC UPC, SC APC, LC UPC, etc.)
Fiber Inspection Guidelines:
• Use the VIAVI P5000i or FiberChek Probe
microscope to inspect both sides of every
connection being used (OTDR Port, Launch
Cable, bulkhead connectors, patch cords, etc.)
• Focus fiber on the screen. If dirty, clean the
end-face.
• If it appears clean, run inspection test.
• If it fails, clean the fiber and re-run inspection
test. Repeat until it passes.
Connect Launch Cable to OTDR port:
1. Inspect the OTDR port on top of the test set.
2. Inspect the fiber end face of Launch Cable.
3. Connect Launch Cable to the OTDR port.
Fiber Optics Software
Figure 1: Equipment Requirements
Figure 2: Inspect Before You Connect (IBYC)
Figure 3: OTDR Port Inspection
VIAVI Solutions

Advertisement

loading

Summary of Contents for Viavi T-BERD/MTS 5800

  • Page 1 Real Time Testing with Expert OTDR This quick card describes how to connect to a fiber under test, configure Expert OTDR settings, run Real Time tests, and analyze results with a VIAVI T-BERD/MTS 5800 and 4100-series OTDR module. Equipment Requirements: •...
  • Page 2 VIAVI Solutions Connect to Fiber Under Test (FUT): The Launch Cable may be connected to the FUT via an optical patch panel (OPP) or an optical coupler: 1. If the interface to the FUT is a patch cord, connect the patch cord to an optical coupler Figure 4: Connecting the Launch Cable to the FUT with a coupler with the same connector type.
  • Page 3 © 2021 VIAVI Solutions Inc. Contact Us +1 844 GO VIAVI To reach the VIAVI office nearest you, Product specifications and descriptions in this (+1 844 468 4284) visit viavisolutions.com/contacts. document are subject to change without notice.