Installation Manual Weighbridge Cable Junction Box PR 6021/68S Translation of the Original Installation Manual 9499 053 16800 Edition 3.8.0 11/22/2021 Minebea Intec GmbH, Meiendorfer Str. 205 A, 22145 Hamburg, Germany Phone: +49.40.67960.303 Fax: +49.40.67960.383...
Page 2
Any information in this document is subject to change without notice and does not represent a commitment on the part of Minebea Intec unless legally prescribed. This product should only be operated/installed by trained and qualified personnel. In correspondence concerning this product, the type, name, and release number/serial number as well as all license numbers relating to the product have to be cited.
WARNING indicates that death or severe, irreversible injury may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. CAUTION Warning of personal injury. CAUTION indicates that minor, reversible injury may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. Minebea Intec EN-3...
NOTICE indicates that damage to property and/or the environment may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. Note: User tips, useful information, and notes. Hotline Phone: +49.40.67960.444 Fax: +49.40.67960.474 eMail: help@minebea-intec.com EN-4 Minebea Intec...
General information Repairs are subject to inspection and must be carried out at Minebea Intec. In case of defect or malfunction, please contact your local Minebea Intec dealer or service center for repair. When returning the device for repair, please include a precise and complete description of the problem.
Storage temperature range -30°C…+80°C Cable screw connections Metal Ex EMC cable screw connections Brass CuZn39Pb3, galv. nickel-plated Insulation impedance (in service temperature range at >1000 MΩ 95% air humidity and U = 500 V) Ex-connection values See Chapter 8. EN-8 Minebea Intec...
S CL III, DIV 1 T4 - 36931-751-07; Type 4X, IP6x CL I, Zone 0, AEx ia IIC T4 - 36931-751-07; Type 4X, IP6x Zone 20, AEx ia IIIC T90°C - 36931-751-07; Type 4X, IP6x T4 Ta= -20°C to 60°C Minebea Intec EN-9...
Page 12
The Ex safety instructions in the appendices must be observed when installing in the Ex area. The relevant protection class must be indi- cated on the rating plate! Note: The marking for ATEX is shown as an ex- ample. Note: See also Chapter 8. EN-10 Minebea Intec...
It is not permissible to connect multiple active intrinsically safe circuits in the junction box. When used in zone 20, the accumulated dust must not exceed a thickness of 5 mm. When used in zone 2, ensure that no voltage peaks more than 40% above the nominal voltage can occur. EN-12 Minebea Intec...
Property damage is possible. If a cable gland is not used, it must be sealed with a supplied locking pin. NOTICE Property damage is possible. Regularly check the fitted cable gland for tightness and re-tighten it, if necessary. EN-14 Minebea Intec...
Connecting cables To connect the junction box to the device, connection cable PR 6135/..or PR 6136/.. (for Ex zone) must be used; max. length: 300 m. Cable connections Note: All components are only shown schematically. Minebea Intec EN-15...
Page 18
Weighbridge Cable Junction Box PR 6021/68S 5 Cable connections Color code black blue green gray white Connection example EN-16 Minebea Intec...
In particular, the entire installation, including the power supply, must be sufficiently protected against lightning. In Ex areas, equipotential bonding must be established to avoid any compensatory currents between the individual conducting system components. Connection example Minebea Intec EN-17...
Electronic corner correction 5.5.1 General information Minebea Intec load cells are produced according to high quality standards and have precisely adjusted output values. Nonetheless, mechanical imbalances can cause impermissible corner load errors to arise, which will need to be offset by soldering in resistors.
Page 21
(in the example 4.68 Ω). This aligns the load cells with the displayed value of the reference cell. Note: The soldering studs are factory short-circuited for the resistances. 3. After that, recalibrate the scale. Minebea Intec EN-19...
Repairs Repairs are subject to inspection and must be carried out at Minebea Intec. In case of defect or malfunction, please contact your local Minebea Intec dealer or service center for repair. When returning the device for repair, please include a precise and complete description of the problem.
We reserve the right not to accept products that are contaminated with hazardous substances (ABC contamination) for repair. Should you have any further questions, please contact your local service representative or our service center. Minebea Intec GmbH Repair center Meiendorfer Strasse 205 A 22145 Hamburg, Germany Phone: +49.40.67960.666...