Cleaning

Set the sensor cleaning setting from SC-U1.
To perform cleaning, press
・ Pressing
Caution
・ Do not press
This may cause the sensor to malfunction.

Manual cleaning

Activate the wiper manually. While cleaning the measurement value of the
target sensor is retained (held).
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "9 Cleaning" and " ◆ Manual Cleaning".

Cleaning interval

Set a time interval to activate the wiper.
This product is a sensor with wipers and it operates as following.
Cleaning Cleaning not performed
1 round
trip of
wiper blade
・ The setting time for the cleaning interval is the time interval between the
end of cleaning and the next cleaning.
・ The cleaning action is fixed at 1 wiper blade round trip per cycle.
Select an item from the following.
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "9 Cleaning" and " ◆ Cleaning Interval".
Item
Cleaning Interval OFF (No cleaning)
30
.
automatically retains the measurement value.
while a sensor with a wiper is in the air.
Cleaning interval
Setting range
10 to 1440 minutes (by 10 minutes)
Cleaning
Cleaning not performed
1 round
trip of
Cleaning interval
wiper blade
Factory default
setting
30 minutes