Measurement item
Main
Turbidity
display
* 度 : Unit is used only in Japan.
The following display resolutions are available.
・ 0.0 - 99.9: 0.1
・ 100 - 499: 1
・ 500 - 3000: 5
The calibration curve for this product is made with formazin
Caution
standard solution.

Response time

Set a signal output response time.
A measurement value is calculated using moving average of the time period
set by the signal output response time.
For example, if the sensor measures values every 5 seconds and the signal
output response time is set to 50 seconds, which is 5 seconds × 10 times,
the moving average of 10 times measurement is calculated and used as the
measurement value.
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.1 Sensor Setting" and " ◆ Response Time".
Response
5 to 600 seconds (by 5 seconds)
Time
The recommended value for this product is 30 seconds or longer.
If the response is too fast with the factory default setting (5 seconds),
change the signal output response time as needed.
14
Display item
Item
Turb.
Turb.
Turb.
Setting range
Factory default
Unit
FTU
Turb. FTU
度 *
No unit
Factory default
setting
5 seconds
setting