ENGLISH VERSION GENERAL INSTALLATION AND MAINTENANCE INSTRUCTIONS HOSPIVAC V3 SIRELLA with PROCOM-2 or CYCLIC 2020 - - - - - - - - - - HOSPIVAC G SIRELLA with PROCOM-2 / CYCLIC 2020 / MILLENIUM RÈGLES DE SÉCURITÉ voir notice n°511927 First apposition of the EC marking: 1998 This document must not be disclosed or reproduced without prior authorisation from MIL’S.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM CONTENTS SECTION 1. INTRODUCTION 1.1..........................3 PECIFICATION PLATE SECTION 2. DESCRIPTION ...
Page 3
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM MEANING OF SYMBOLS USED IN THESE INSTRUCTIONS Symbols are used in these instructions to help you understand them and to indicate particular points which you need to act upon or consider.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 1. INTRODUCTION We strongly advise you to read these installation instructions, before first switching on your vacuum unit. The damage that could be caused by setting the machine up or installing it incorrectly would not be covered by our guarantee.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 3. OPERATION The air conveyed by the vacuum network passes through the main bacteria filter, enters the main receiver and is then sucked by the vacuum pump(s) of the vacuum unit/plant.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM For information purposes, the table below gives the amounts of heat that must be extracted, according to the installed capacity in kW.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 6. OUTLINE DIAGRAM 6.1. Mechanical architecture HOSPIVAC V3 SIRELLA régulation vide stable Constant vacuum regulation Vacuostat de sécurité...
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 6.2. Electrical architecture . HOSPIVAC G2 / G3 MIL Depth : 270 MILLENIUM control system Pump temperature control...
Page 9
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM Electrical cabinet - wall mounted fixation 4 x Ø9 .HOSPIVAC G4 / G5 / G6 PM depth : 305...
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 7. INSTALLATIONS 7.1. For HOSPIVAC V3 SIRELLA On the framework assembly: The three claw vacuum pumps, ...
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 7.2. For HOSPIVAC G ØD Suction Exhaust Hot air outlet Cooling air 4 holes M10 ØD ØE F 1"1/2 M 1"1/2...
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 7.3. Power supply (see electrical circuit diagram attached) Power supply must be removed prior to any intervention on the station Connection to the power supply must be carried out by the exploiting.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM The wire sections are given for information to 5A / mm Even so, the sections shall be defined by taking contacts with the wire provider.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 8. COMMISSIONING HOSPIVAC systems are used only for suction of non-aggressive gas, with less than 22% of oxygen gas.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 9. SETTINGS 9.1. Setting pressure Qualified and authorised personnel should only carry out any modification to the settings.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 9.1.2. Pressure setting for PROCOM-2, CYCLIC 2020 and MILLENIUM For settings procedure, see the attached regulation manual. Network sensor C2V...
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 10. ALARMS 10.1. For the MILLENIUM Alarms appear on the display and are: Disjunction, Network pressure (option).
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 11. CLEANING AND MAINTENANCE IMPORTANT: For this operation, use all necessary protective measures (gloves, mask, overalls etc.) then destroy contaminated consumables (antibacterial cartridges, condensates etc.) in according...
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 11.1. Filtration During and after maintenance, respect basic hygiene rules. Before ordering tools used for maintenance, clean them with a disinfectant.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 11.4. Changing the hose with vacuum applied To change the hose, proceed methodically in order to avoid interrupting vacuum production and always remember to isolate the unit from the rest of the network.
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 14. ELECTROMAGNETIC COMPATIBILITY The HOSPIVAC units with PROCOM2, C2020 or MILLENIUM automate needs special precautions regarding EMC.
Page 22
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM Guidance and manufacturer’s declaration – electromagnetic immunity - for ME equipment and ME systems that are not life-supporting The HOSPIVAC units with PROCOM2, C2020 or MILLENIUM automate is intended for use in the electromagnetic environment specified below.
Page 23
General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM The HOSPIVAC units with PROCOM2, C2020 or MILLENIUM automate can be affected by communication portables or mobiles RF’s equipment.