MIL'S HOSPIVAC V3 SIRELLA General Installation And Maintenance Instructions

MIL'S HOSPIVAC V3 SIRELLA General Installation And Maintenance Instructions

Medical vacuum pump with procom-2, cyclic 2020, or with procom-2 / cyclic 2020 / millenium

Advertisement

-
ENGLISH VERSION
-
GENERAL INSTALLATION AND MAINTENANCE INSTRUCTIONS
HOSPIVAC V3 SIRELLA
with PROCOM-2 or CYCLIC 2020
- - - - - - - - - -
HOSPIVAC G SIRELLA
with PROCOM-2 / CYCLIC 2020 / MILLENIUM
RÈGLES DE SÉCURITÉ voir notice n°511927
First apposition of the EC marking: 1998
This document must not be disclosed or reproduced without prior authorisation from MIL'S.
Failure to respect this requirement may lead to prosecution.
-
11/16 – MANUAL N° 519923 VERSION 4
-

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HOSPIVAC V3 SIRELLA and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for MIL'S HOSPIVAC V3 SIRELLA

  • Page 1 ENGLISH VERSION GENERAL INSTALLATION AND MAINTENANCE INSTRUCTIONS HOSPIVAC V3 SIRELLA with PROCOM-2 or CYCLIC 2020 - - - - - - - - - - HOSPIVAC G SIRELLA with PROCOM-2 / CYCLIC 2020 / MILLENIUM RÈGLES DE SÉCURITÉ voir notice n°511927 First apposition of the EC marking: 1998 This document must not be disclosed or reproduced without prior authorisation from MIL’S.
  • Page 2: Table Of Contents

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM CONTENTS     SECTION 1. INTRODUCTION     1.1..........................3 PECIFICATION PLATE     SECTION 2. DESCRIPTION  ...
  • Page 3 General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM MEANING OF SYMBOLS USED IN THESE INSTRUCTIONS Symbols are used in these instructions to help you understand them and to indicate particular points which you need to act upon or consider.
  • Page 4: Introduction

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 1. INTRODUCTION We strongly advise you to read these installation instructions, before first switching on your vacuum unit. The damage that could be caused by setting the machine up or installing it incorrectly would not be covered by our guarantee.
  • Page 5: Operation

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 3. OPERATION The air conveyed by the vacuum network passes through the main bacteria filter, enters the main receiver and is then sucked by the vacuum pump(s) of the vacuum unit/plant.
  • Page 6: Regulation By Sensor

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM For information purposes, the table below gives the amounts of heat that must be extracted, according to the installed capacity in kW.
  • Page 7: Outline Diagram

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 6. OUTLINE DIAGRAM 6.1. Mechanical architecture HOSPIVAC V3 SIRELLA régulation vide stable Constant vacuum regulation Vacuostat de sécurité...
  • Page 8: Electrical Architecture

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 6.2. Electrical architecture . HOSPIVAC G2 / G3 MIL Depth : 270 MILLENIUM control system Pump temperature control...
  • Page 9 General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM Electrical cabinet - wall mounted fixation 4 x Ø9 .HOSPIVAC G4 / G5 / G6 PM depth : 305...
  • Page 10: Installations

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 7. INSTALLATIONS 7.1. For HOSPIVAC V3 SIRELLA  On the framework assembly:  The three claw vacuum pumps, ...
  • Page 11: For Hospivac G

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 7.2. For HOSPIVAC G ØD Suction Exhaust Hot air outlet Cooling air 4 holes M10 ØD ØE F 1"1/2 M 1"1/2...
  • Page 12: Power Supply (See Electrical Circuit Diagram Attached)

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 7.3. Power supply (see electrical circuit diagram attached) Power supply must be removed prior to any intervention on the station Connection to the power supply must be carried out by the exploiting.
  • Page 13: Connections To The Pneumatic Network

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM   The wire sections are given for information to 5A / mm Even so, the sections shall be defined by taking contacts with the wire provider.
  • Page 14: Commissioning

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 8. COMMISSIONING HOSPIVAC systems are used only for suction of non-aggressive gas, with less than 22% of oxygen gas.
  • Page 15: Settings

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 9. SETTINGS 9.1. Setting pressure Qualified and authorised personnel should only carry out any modification to the settings.
  • Page 16: Flexo Time Duration

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 9.1.2. Pressure setting for PROCOM-2, CYCLIC 2020 and MILLENIUM For settings procedure, see the attached regulation manual. Network sensor C2V...
  • Page 17: Alarms

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 10. ALARMS 10.1. For the MILLENIUM Alarms appear on the display and are:  Disjunction,  Network pressure (option).
  • Page 18: Cleaning And Maintenance

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 11. CLEANING AND MAINTENANCE IMPORTANT: For this operation, use all necessary protective measures (gloves, mask, overalls etc.) then destroy contaminated consumables (antibacterial cartridges, condensates etc.) in according...
  • Page 19: Filtration

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 11.1. Filtration During and after maintenance, respect basic hygiene rules. Before ordering tools used for maintenance, clean them with a disinfectant.
  • Page 20: Changing The Hose With Vacuum Applied

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM 11.4. Changing the hose with vacuum applied To change the hose, proceed methodically in order to avoid interrupting vacuum production and always remember to isolate the unit from the rest of the network.
  • Page 21: Electromagnetic Compatibility

    General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM SECTION 14. ELECTROMAGNETIC COMPATIBILITY The HOSPIVAC units with PROCOM2, C2020 or MILLENIUM automate needs special precautions regarding EMC.
  • Page 22 General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM Guidance and manufacturer’s declaration – electromagnetic immunity - for ME equipment and ME systems that are not life-supporting The HOSPIVAC units with PROCOM2, C2020 or MILLENIUM automate is intended for use in the electromagnetic environment specified below.
  • Page 23 General installation and maintenance instructions HOSPIVAC V3 SIRELLA with PM-2 or CYCLIC 2020 HOSPIVAC G SIRELLA with PM-2 or C2020 or MILLENIUM The HOSPIVAC units with PROCOM2, C2020 or MILLENIUM automate can be affected by communication portables or mobiles RF’s equipment.