User Manual
SPI-P170
Hydrogen Peroxide

Advertisement

loading
Need help?

Need help?

Do you have a question about the SPI-P170 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SPI SPI-P170

  • Page 1 User Manual SPI-P170 Hydrogen Peroxide...
  • Page 3 SPI – P 170 P E R O X I D E Index SPI-P170 ........................... 1 1. INTRODUCTION ........................3 2. DESCRIPTION AND OPERATION ....................5 3. SAFETY ............................. 7 4. TRANSPORT AND STORAGE ....................... 8 5. ASSEMBLY AND INSTALLATION ....................9 6.
  • Page 4 The manual for the SPI-P170 is meant for the following authorized employees: Electrotechnical staff Watertechnical staff Laboratory staff This manual is made for the installation and operation of the SPI-P170 Hydrogen Peroxide. In this manual you will find various enumeration characters: Enumeration of functions (1.)
  • Page 5 SPI – P 170 P E R O X I D E 1. Introduction Purpose of the SPI-P170 The SPI-P170 is designed for correctly measuring, controlling, and guarding a water treatment process. The SPI-P170 is suitable for the following sectors: ...
  • Page 6 No higher supply voltage than 12VDC is used. Background information After years of collaboration with a team of specialists in the field of water treatment, the SPI measure and control system was introduced. The SPI-P170 is the fourth generation of its kind. The SPI-P170 is a water control system based on the colorimetric measuring method that utilizes peroxide reagent discoloration.
  • Page 7: Description And Operation

    When you have the optional SPI-REMOTE software, it is possible control and see the information of the SPI using the network or internet. With the optional SPI-GRAPHS software, you can use the data downloaded with the remote software, and use this for data acquisition.
  • Page 8 SPI – P 170 P E R O X I D E Analysis-unit Control-unit Figure 2.1.1. Buffer jar Reagent vial Filling tube Drain tap Supply valve Reagent valve Overflow tube Drain valve Filter Figure. 2.2.1. V1 20130904...
  • Page 9 E R O X I D E 3. Safety Safety regulations The SPI-P170 is as careful as possible designed with the eye on safety. We tried to reduce the safety risks to a minimum.  The SPI-unit is supplied with a 12VDC adapter via a power outlet.
  • Page 10: Transport And Storage

    11. Take the SPI-unit off the wall. Transport 1. After disassembly, make sure the SPI-unit is placed in a firm box, with the components facing upwards. This may also be standing upward. 2. Protect the corners well from impacts.
  • Page 11 The power outlet is in an acceptable range of the SPI-unit (100-240VAC 47-63Hz)  The return to the buffer tank is taken care of and brought to where the SPI is going to be  The drain to the sewer is taken care of and brought to where the SPI is going to be Please follow the instructions for installing the SPI-P170: 1.
  • Page 12 SPI – P 170 P E R O X I D E Buffer jar Control-unit Supply and drain Figure 5.1.1. Figure 5.2.1. 6. Commissioning V1 20130904...
  • Page 13: Operation

    15.4. Indication LEDs After the SPI-P170 has configured itself, the blue LED will start flashing. On the front panel of the SPI-P170 control-unit are a total of 3 LEDs. The explanation of these LEDs are as following:  Blue ON/OFF ;...
  • Page 14 SPI – P 170 P E R O X I D E Using the keyboard, you can perform all operations. You simply follow the options through the menu structure on the display. Navigating through the menu structure is done using the arrow buttons. The display displays a maximum of 4 lines a time.
  • Page 15: Adjust V Alue

    SPI – P 170 P E R O X I D E select this symbol with the cursor and you press enter, you will go into this menu. Going back is always done with the left arrow (). See figure 7.2.1.
  • Page 16 Using the main menu, all important functions of the SPI-P170 can be reached. The main menu is what you get to see after starting the SPI-P170, and is where you always get back to. The main menu consists of the following options: ...
  • Page 17: Cel L Fa Ilure

    SPI – P 170 P E R O X I D E 9. Overview You can find the most important measurements and values and relevant information in the overview menu as in figure 9.1. You can enter this menu by selecting Overview from the main menu.
  • Page 18 E R O X I D E 10.Calibration All measurements connected to the SPI-P170 must be regularly checked and, if necessary, be calibrated. This ensures the quality of the measurements made. You can enter this menu by selecting Calibration from the main menu.
  • Page 19 SPI – P 170 P E R O X I D E 10.1.1 Calibrating with handheld When you select Calibrating with handheld the menu as in figure 10.1.1.1 will appear. Here you can perform a calibration using a third party handheld. Now perform the following instructions: 1.
  • Page 20 SPI – P 170 P E R O X I D E 10.1.2 Calibration with fluids Calibrating using fluids guarantees a more dependable calibration then when using a handheld. The system lets you first perform a zero point measurement, and after that a measurement using a control fluid of your choice.
  • Page 21 SPI – P 170 P E R O X I D E  6. If the value is correct, select and press enter ( ) on Begin 52 ppm calib. The menu of figure 10.1.2.3 now appears. Cal at 52 ppm...
  • Page 22 SPI – P 170 P E R O X I D E 10.1.3 Reset Calibration If you are not satisfied with the current Hydrogen Peroxide calibration or you don’t have the right tools to calibrate, you can reset the calibration settings to the factory settings. To do a calibration reset, select Restore calibration from the Hydrogen Peroxide calibration menu.
  • Page 23 2. Take a new vial and place it in the designated holder on top of the analysis-unit. 3. Replace the standard cap of the vial and replace this with the cap supplied with the SPI-P170. This cap contains 2 holes; one for the tube, the other for aeration.
  • Page 24 SPI – P 170 P E R O X I D E 5. By now the menu as in figure 10.1.4.2.2 has appeared. The cell is prepared. Put you r finger on top of tube. Pre ss ( V) Figure 10.1.4.2.2.
  • Page 25 Cycle time (The time the SPI-P170 will ‘wait’ between 2 measurements in seconds. Setting this at 300, means that the SPI will wait 5 minutes, before the next measurement starts. If the color valve time is set at 0, the automatic peroxide reagent dosing is activated. This means that the dosing of a shot of peroxide color will be adapted to the use of the reagent.
  • Page 26 More time means fewer measurements, but it saves more peroxide color. The SPI-P170 will temporary stop measuring and controlling as long as this menu is opened. The pumps will remain pumping at the same level as before this menu was opened.
  • Page 27 SPI – P 170 P E R O X I D E When you select pH from the calibration menu, the menu as in figure 10.2.1 will appear. In this menu it is possible to calibrate the pH. For calibrating the pH value, it is important to have the right tools at hand. The pH is calibrated using 2 independent constant buffer solutions with different pH values.
  • Page 28 SPI – P 170 P E R O X I D E 7. Check the value at Actual pH. Wait for this value to get stable, then confirm  with ( 8. Now flush the electrode with clean water. 9. Execute steps 2 -8 with the pH 4.01 calibration (or your own calibration).
  • Page 29 E R O X I D E 11 Settings In the settings menu it is possible to change the settings of specific parts of the SPI-P170. You can enter this menu by selecting Settings from the main menu. You can choose between the following submenus: ...
  • Page 30 SPI – P 170 P E R O X I D E When you chose for Hydrogen settings in the previous menu, you’ll see the screen as in figure 11.1.1. Here you can view or change the following settings: ...
  • Page 31  Delta (this setting controls the maximum and minimum pump setting. The SPI-P170 calculates the average pump control. The pump cannot go over xx% of this average value) ...
  • Page 32 The pump cannot go over xx% of this average value)  Prop factor (his setting controls the proportional gain of the pH controller system. The SPI- P170 constantly calculates the difference between the setpoint and the actual pH value. The more difference, the faster or slower the pump will work.
  • Page 33 SPI – P 170 P E R O X I D E  Max pulse time (the maximum time at which the pump can pump on maximum setting. If after this time the pump is fully pumping, and the measured value still...
  • Page 34 SPI – P 170 P E R O X I D E When you chose for Flow settings in the settings menu, you’ll see the screen as in figure 11.3.1. Here you can view or change the following settings: ...
  • Page 35 Year (the years of the current date) It could happen that the SPI-P170 has reset the date & time settings at a power failure. If this is the case, the SPI will start counting from it’s default date and time values. The cause of this is most likely a bad battery.
  • Page 36 SPI – P 170 P E R O X I D E Day hou r Day min ute Figure 11.5.1. Nig ht h our Nig ht minute These settings make different settings for day/night times of the system possible. For example, at night, you could set a lower value for the Hydrogen Peroxide setpoint You can find these settings in the Hydrogen Peroxide or pH settings as explained in chapter 11.1 and 11.2.
  • Page 37 All current alarms and their alarm values will be shown in the Alarms menu. If an alarm has presented itself, the red LED will light on the keypad of the SPI-P170. You can enter this menu by selecting Alarms from the main menu.
  • Page 38 In this case, the flow protection is turned off. More about this in chapter 13. Not present or maintenance mode (X) This character indicates the measurement is not present on the SPI and has been turned off. It can also mean the user is making maintenance on this part.
  • Page 39 12.2 Alarm contact The SPI-P170 contains an alarm contact. This is a potential free contact that can be used to transmit the alarm signals to equipment of third parties. The alarm contact is closed when an alarm is being detected on the SPI.
  • Page 40 SPI – P 170 P E R O X I D E 12.3 Solving alarms With correct use and regular maintenance, the least hardware errors will occur. For maintenance, see chapter 15. For solving problems you follow the following steps: 1.
  • Page 41 SPI – P 170 P E R O X I D E Example On the keypad, the red LED is lighting up. The alarm “A09: pH PUMP” is shown on the Alarm menu. This alarm means that the pump has been at full power for over a course of <1800> seconds (this time is adjustable), and the pH value is still high.
  • Page 42 The temporarily turning off of the SPI-P170 with data preservation, is called the hot start As the unit is commissioned for a longer period of time, the battery ages. This is causing the SPI no longer retaining the clock settings. As soon as this seems to be the fact, we advise you to take steps to replace the battery.
  • Page 43: Manual Operation

    E R O X I D E 13.Manual operation The SPI-P170 contains the possibility to let the user control the Hydrogen Peroxide or the acid pumps on manual or semi-automatic mode. When you select Manual operation from the main menu, the screen as figure 13.1 appears.
  • Page 44 SPI – P 170 P E R O X I D E Setting mode The set the right mode for a pump, you select the mode of that specific pump, and press enter  ). You can fill in the following values for each mode: 1.
  • Page 45 SPI – P 170 P E R O X I D E 14.Reports When you select Reports from the main menu the screen as figure 14.1 appears. In this menu you can read back and check various reports such as: 1.
  • Page 46 SPI – P 170 P E R O X I D E When you select Calibration reports from the previous menu, the screen as in figure 14.2.1 appears. In this screen you can see all logged calibrations. The first log you see is the most recent.
  • Page 47 SPI – P 170 P E R O X I D E When you select Maintenance reports from the previous menu, the screen as in figure 14.3.1 appears. In this screen you can see all logged maintenance services. The first log you see is the most recent.
  • Page 48: Maintenance

    Check the inside of the SPI-unit for possible water or moisture Checking the valves on correct operation One of the most important parts of the SPI-P170 are the valves that regulate the supply and drain of measuring water and chemicals.
  • Page 49 Every once in a while we advise to check the measurements for abnormalities, and when necessary, to calibrate them. When this is the case, calibrate the measurements of the SPI-P170 with well calibrated equipment of third parties. More about calibrating is described in chapter 10.
  • Page 50 16. Configuration When you select Configuration from the main menu, the screen as figure 16.1 appears. With the help of this menu you can check and change various settings used for the configuration of your SPI system. The menu contains the following options: 1.
  • Page 51 5. Acid pump p/m (Fill in the maximum amount of pulses of your acid pump) 6. Extern contact (Set if you want to be able to connect the SPI to the internet (1) or not (0)) 7. Log period (Set the time in seconds the SPI system waits before creating a new log report)
  • Page 52 13. Cell alarm delay (Set the time in seconds the system delays the alarm of starting and complete pollution) 14. System ID (Set a custom identification number to this SPI for recognition. The numbers 1 – 254 can be used) 15.
  • Page 53 When you chose for Version info in the Configuration menu, the screen as in figure 16.2.1 appears. In the version information menu it is possible to view the following information: 1. Software (the current version of the SPI-P170 Hydrogen Peroxide software) 2. System ID (Your custom system ID of this SPI model) You can use this information if the supplier asks for this for technical assistance.
  • Page 54 Because of the discoloration, the water will get more dark, and absorb more light. When this happens, less light will fall on the receiver, resulting in less voltage. The SPI will measure this change. A measurement without peroxide color should give an internal value of around 1000. A good measurement with peroxide color added will be much lower.
  • Page 55 SPI – P 170 P E R O X I D E 16.3.2 Calibrating zero value When the zero value starts to differ a lot from the normal value of 1000, it is important to check if the cell is clean. If the cell is clean, the cell might need to be calibrated. Calibrating implies that the zero value must be set back to its default value of 1000.
  • Page 56 Figure 16.4.1. 16.5 System reset The SPI-P170 contains an option to reset the system using the screen instead of taking out the adapter out of the socket. This way, you can reset the SPI using the SPI-REMOTE software. When you ...
  • Page 57 E R O X I D E Decommissioning If you’d like to decommission the SPI-P170 for (longer) periods, follow the below procedure: 1. Remove the adapter from the power outlet. 2. Remove the peroxide color vial from the holder an rinse the tube with clean water. (please take the necessary safety precautions).
  • Page 58 SPI – P 170 P E R O X I D E 18. Discarding The SPI-P170 contains electronic components. Inform to the possibilities to separate and recycle these components during discarding. Remove possible chemical remains like Hydrogen Peroxide or the corresponding peroxide color and discard these as chemical waste.
  • Page 59 SPI – P 170 P E R O X I D E Notes V1 20130904...