Agilent Technologies 6000 Series User Manual page 60

Oscilloscopes
Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

2
Testing Performance
60
Equipment Required to Verify Horizontal ∆t Accuracy
Table 11
Equipment
Critical Specifications
Oscilloscope
Stability 5 ppm after 1/2 hour
Calibrator
Cable
BNC, 3 feet
Ω BNC (f) to BNC (m)
Feedthrough
50
1 Connect the equipment:
a For 300 MHz, 500 MHz, and 1 GHz oscilloscope models,
connect the calibrator output to the oscilloscope channel 1
input.
b For 100 MHz oscilloscopes, use a 50Ω feedthrough
termination to connect the calibrator output to the
oscilloscope channel 1 input.
2 Set up the signal source.
1 Select Marker on the oscilloscope calibrator.
2 Set the calibrator for 100 µs markers (period = 100 µs).
3 Set up the oscilloscope.
a Set channel 1 Coupling to DC.
b On 1 GHz, 500 MHz and 300 MHz models only, set channel
1 Imped to 50 Ohm.
c Press the Display key, then set the Vectors softkey to off.
d Press the AutoScale key.
e Set the time base to 20 µs/div.
f Press the Main/Delayed key, then set the Time Ref softkey to
Left.
g Adjust the Trigger Level knob to obtain a stable display.
4 Press the Quick Meas softkey, set the Source softkey to 1, then
press Select and choose Period. Press the Measure softkey and
measure the following:
Period 100 µs — The test limits are 99.8 µs to 100.2 µs.
Recommended
Model/Part
Fluke 5820A
Agilent 10503A
Agilent 0960-0301