Service Guide
Table 9
Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting
Oscilloscope
Calibrator Setting
5 V/Div
35 V
2 V/Div
14 V
1 V/Div
7 V
500 mV/Div
3.5 V
200 mV/Div
1.4 V
100 mV/Div
700 mV
50 mV/Div
350 mV
20 mV/Div
140 mV
10 mV/Div
70 mV
5 mV/Div
35 mV
1 (for 1 GHz, 500 MHz, and 300 MHz models)
2 mV/Div
14 mV
2 mV/Div
(for 100 MHz models)
14 mV
2
1 mV/Div
7 mV
1
Full scale is defined as 32 mV on the 2 mV/div range for 300 MHz, 500 MHz, and 1 GHz models.
2
Full scale is defined as 16 mV on the 1 mV/div range for 100 MHz models.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
c Press the Acquire key. Then press the Averaging softkey and
set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2 Press the Cursors key, set the Mode softkey to Normal, then
press the X Y softkey and select Y. Press the Y1 softkey, then
use the Entry knob (labeled
the Y1 cursor on the baseline of the signal.
Testing Performance
Test Limits
34.04 V
to
35.96 V
13.616 V
to
14.384 V
6.808 V
to
7.192 V
3.404 V
to
3.596 V
1.3616 V
to
1.4384 V
680.8 mV
to
719.2 mV
340.4 mV
to
359.6 mV
136.16 mV
to
143.84 mV
68.08 mV
to
71.92 mV
34.04 mV
to
35.96 mV
13.232 mV
to
14.768 mV
13.616 mV
to
14.384 mV
6.616 mV
to
7.384 mV
on the front panel) to set
2
51