Replacing 6000A Assemblies
108
7 To reinstall the front panel:
a Align the hooks on top of the front panel with their
connection holes in the sheet metal and display mount.
b Swing the front panel down and ensure that the power
switch, intensity shaft and BNC connectors are aligned
with the holes in the front panel.
c Push the front panel until the two retainer tabs click into
place in the deck.
d Reinstall the T6 screws on the BNC connectors.
e Connect the keyboard ribbon cable.