9. Error Messages
9-1. Notifications
Notifications
The system check was failed or has
an error.
Please check the results again.
Contact the system administrator.
The QC was failed or has an error.
Please check the results again.
The temperature is not within
operational range.
(Current temperature: 38℃ / (100.4℉)
The admin account automatically
logged out.
Please log in first.
Please input the master password.
Incorrect password.
Please input the master password
again.
Description and Measure
System Check is not performed.
Displayed when system check is not performed.
[Setting screen → System Check → Run System
Check]
QC not performed.
Displayed when touching QC! icon on the home
screen in the case of when the QC Test has not been
performed.
QC not performed.
Displayed when touching QC! icon on the home
screen in the case of when the QC Test has not been
performed.
Administrator account logout
Displayed when administrator account logs out
before power off or reboot.
Attempted access in a logged off state
Displayed when the setting menu requires
administrator account login.
Master password is not inputted.
Displayed when the 'DONE' button is touched without
master password input.
→ Enter the master password and press OK.
Password input error
Displayed when wrong password is inputted.