6. Running a QC Test (Control Solution Test)
STEP 7. Check the result of QC Test
When test is completed, result of QC test is displayed on the same screen.
a.
Check the result and remove the cartridge from the tray.
b.
Touch the 'DONE' button → Return to the home screen.
c.
QC Test Result
Positive, Negative test All Pass :
Normal operation of test cartridge
One or more of the Positive, Negative test Fail :
Accuracy of the actual test is not verified.
Conduct the test again or contact the administrator.
STEP 8. Discard the used test cartridge.
Remove the cartridge from the tray referring to the figure below and place it in a safe
a.
place
27
.